Installation of New
Reference Electrode
Reference Electrode
with Filling Solution
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
1 . . Unpack .the .reference .electrode .(Cat . .No . .210058) .and .its .instruction .
sheet .from .the .shipping .box . . .
2 . . Carefully .remove .protective .caps .from .bottom .and .sidearm .of .electrode . .
Save .caps .for .future .storage .
3 . . Shake .out .as .much .of .the .fill .solution .through .the .fill .hole .as .possible . .
Drain .through .the .sidearm .or .use .a .pipette .or .syringe .if .needed .
4 . . While .passing .1/8 .inch .tubing .into .electrode .sidearm, .slide .1⁄4 .inch .
tubing .over .sidearm . . . T he .outside .tubing .should .extend .3/8- .1⁄2 .inches .
over .sidearm . . .Refer .to .Figure II-13 .
Reference Electrode
Tubing Bottle Adaptor
Small 1/8" Tubing Inside
Figure II-3
5 . . Remove .cap .and .fluid .seal .from .reference .electrode .filling .solution .
bottle .(Cat . .No . .150072) . . .Hold .bottle .in .upright .position . . .Check .
that .rubber .gasket .is .properly .aligned, .then .connect .cap .end .of .tubing .
assembly .to .bottle . . . T he .1/8-inch .tubing .should .extend .into .the .bottle .
6 . . Hold .reservoir .bottle .above .electrode .with .bottle .cap .end .down . .
Electrode .should .be .horizontal .with .sidearm .pointing .up . . .Gently .shake .
electrode .to .allow .any .trapped .air .bubbles .to .rise .into .bottle .as .electrode .
fills .with .solution .
7 . . Dry .off .ceramic .frit .on .base .of .the .electrode .with .tissue .paper . . .Squeeze .
bottle .for .a .few .seconds . . .A .small .amount .of .filling .solution .should .
bead .up .on .frit .surface, .indicating .good .fill .solution .flow . . .If .no .
moisture .is .visible .electrode .is .clogged .and .should .be .replaced .
Filling Solution
Gasket
Bottle Cap
Large 1/4" Tubing
上海肯强贸易有限公司
http://www.kq17.com.cn
Instrument Preparation
Electrode Sidearm
ROSS
Coil
Ceramic Frit
II-4