Take standard precautions for transmission of blood borne pathogens
to avoid contamination .
The Reader should be cleaned between patients .
If a Sensor breaks inside a patient, remove with tweezers, treat any
medical complications and call Customer Service .
Use of the System is not recommended in the critically ill population
since performance is unknown due to different conditions and
medications .
Sensor placement is not approved for sites other than the back of the
arm . If placed in other areas, the Sensor may not function properly .
If the Sensor Kit package or contents or the Reader appear to be
damaged, do not use as there may be a risk of electric shock, no results,
and/or infection .
Store the Sensor Kit between 39°F-77°F . While you don't need to keep
the Sensor Kit in a refrigerator, you can as long as the refrigerator is
between 39°F-77°F .
Store the Sensor Kit between 10-90% non-condensing humidity .
The System does not provide real-time results . Patients need to rely on
blood glucose readings for monitoring glucose during System use .
Clean hands prior to Sensor handling/insertion to help prevent
infection .
Clean the application site and ensure that it is dry prior to Sensor
insertion . This helps the Sensor stay attached to the body .