Using the Sensor
After you assemble and apply the Sensor to your patient's body, start
the Sensor with the Reader and confirm it is working . The Sensor stores
glucose readings every 15 minutes for up to 14 days . The first reading is
stored 1 hour after the Sensor is successfully started .
CAUTION:
The Sensor Pack and Sensor Applicator are packaged as a set
(separately from the Reader) and have the same Sensor code .
Check that the Sensor codes match before using
the Sensor Pack and Sensor Applicator . Sensor
Packs and Sensor Applicators with the same
Sensor code should be used together or
Sensor glucose readings may be incorrect .
14
FreeStyle Libre
Sensor Pack
CODE
XXX
XXXXXXXXX
YYYY-MM-DD