Summary of Contents for Chauvin Arnoux C.A 895

  • Page 1 C.A 895 Détecteur de CO CO detector CO-Detektor Rilevatore di CO Detector de CO F R A N Ç A I S Notice de fonctionnement E N G L I S H User's manual D E U T S C H...
  • Page 2: Table Of Contents

    Significance of the symbol CAUTION! Please consult the Operating Instructions before using the device. In these Operating Instructions, failure to follow or carry out instructions preceded by this symbol may result in personal injury or damage to the device and the installations. Definition of CAT III This voltage surge category III clamp complies with stringent reliability and availability requirements, corresponding to fixed industrial and...
  • Page 3: Presentation

    1. PRESENTATION The CA895 detector measures the level of carbon monoxide present in the surrounding atmosphere from 0 to 1,000 ppm. For this purpose it uses a catalytic chemical sensor which does not consume any chemical substance. This portable instrument is particularly suitable for detecting the levels of CO potentially dangerous for human beings in heating, refrigeration, ventilation/air-conditioning installations and, in general, in industrial, commercial or residential environments where combustion gas may...
  • Page 4: Precautions For Use

    • the gas or fumes given off by combustion are not too diluted in the air before reaching the CO sensor of the measuring instrument the CO emission may actually be very localised and you can measure strong concentrations at one measuring point and a weak concentration only a few centimetres from your first measuring point.
  • Page 5: Procedure

    symbols maximum measurement value Hold last value memorised Low battery 3-way switch: OFF: instrument switched off Normal: short response time measurement Average: integrated measurement (stabilised) on / off push button for backlighting HOLD on / off push button to activate the HOLD function on / off push button to activate the MAX function protective cap for the tester calibration screw 4.
  • Page 6: Sensor Calibration

    4.2 Calibration of the sensor (zero reset) Wait until the sensor has been stable for at least 30 seconds. Preferably place the sensor outside in a CO-free zone, remove the CAL screw protective cap and reset the display to 0 +- 1 ppm using the screwdriver provided.
  • Page 7: To Order

    Agent in your country. Repair • Repairs under or out of guarantee: please return the product to your distributor. 7. TO ORDER C.A 895…………………………………………..………..P01. 6510.01Z Supplied with a shockproof sheath, a 9V battery and these operating instructions. Spare parts: Pile 9 V…………………………………………....………..P01. 1007.32...
  • Page 8: Appendix

    8. APPENDIX...