Cal.lock() - Keithley DMM7510 Calibration Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 4: TSP command reference

cal.lock()

This function prevents access to instrument calibration.
Type
TSP-Link accessible
Function
Yes
Usage
cal.lock()
Details
Calibration data is locked during normal operation. To perform calibration, you must unlock
calibration.
This command does not save calibration data.
Calibration data is lost if it is you do not save it before sending cal.lock(). Use cal.save() to
save the data.
An error is generated if this command is issued when calibration is already locked.
Example
cal.unlock("KI000CAL")
-- Perform operations to generate the calibration data
cal.save()
cal.lock()
Unlock the calibration for the instrument using the default password.
Save the calibration data.
Lock the calibration data.
Also see
cal.save()
(on page 4-18)
cal.unlock()
(on page 4-19)
4-16
Affected by
Where saved
Model DMM7510 Calibration and Adjustment Manual
Default value
DMM7510-905-01 Rev. B / February 2016