NOTICE
1. If the red LED (2) is continuously lit, the battery is dis-
charged and requires immediate charging. The pump
motor is also deactivated. In this case the foot pump
and the auxiliary hand pendant must be used until AC
power is restored or the batteries become sufficiently
charged.
2. If the green LED (4) is lit, the batteries are fully charged
and the OR table can be used normally without AC
power.
¡
The following prerequisites must be met:
¡ The OR table is turned on.
¡ Sufficient charge on batteries (if unplugged).
Locking the OR table
¡ Press FLOOR LOCK:
NOTICE
1. Unlocking with two presses of the button is required for
all D860, D830, and D770 tables. D850, D820 and D760
tables with serial numbers prior to those specified below
require only one press of the button to unlock: 35-XXXXX-
T-10431, 34-XXXXX-T-10167, 33-XXXXX-T-10120. Additional-
ly, all D850 tables starting with 3260-, 3261-, 3262-, 3263;
and D820 tables starting with 3282, require 1 press of the
button to unlock.
¡ The OR table is now locked.
Unlocking the OR table
¡ Press FLOOR LOCK (2x in quick succession).