Recommended separation distances between portable and mobile RF
communications equipment and this device
This device is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of this device can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
this device as recommended below, according to the maximum output power of the communications equipment
Rated maximum
output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Separation distance according to frequency of transmitter m
150 kHz to 80
80 MHz to 800 MHz
d
1
2 .
MHz
=
d
1
2 .
P
=
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
800 MHz to 2,5 GHz
P
d
2
3 .
P
=
0.23
0.73
2.3
7.3
23
22