Apex Medical iCH Prime Instruction Manual page 18

Cpap system
Table of Contents

Advertisement

English
English
Condensation in mask or
flexible tube
Water Leakage
Error / Warning Messages shown on LCD
M e s s a g e t yp e
Error:
Primary function can't
execute.
Warning:
NOTE: When the warning message appears, contact your doctor or equipment provider to
download the memory data and reset the meter.
1. The heater plate setting is
too high.
2. The operating environment
or position of heated
humidifier is not correct. The
temperature close to mask
or flexible tube is low.
1. Wwater chamber is not
properly assembled.
2. Water chamber worn out.
D e f i n i t i o n
Error for abnormal system settings
Error for flow generator failure
Error for abnormal timer setting or timer
failure
Error for flow sensor failure
Error for heater failure (thermal fuse
blows off)
Out of system memory
System memory is nearly full
SD card module communication failure
Remove the SD card while data is being
processed
SD card is full
1. Adjust the control knob to low
temperature setting.
2. Remove any air conditioner which
may nearby the heated humidifier. Or
keep room temperature up near 25°C
1. Remove the water chamber from the
heated humidifier, pour out the water
and reassemble the water chamber
again, making sure water chamber
closes securely, and fill the water till it
reaches fill line and check if it still
leaks or not.
2. Replace with a new water chamber.
16
M e s s a g e i n L C D
Error 001
Error 002
Error 003
Error 004
Error 005
Warn 001
Warn 002
Warn 003
Warn 004
Warn 005