Advertisement

Quick Links

GB - User's manual
D38N
北京海洋兴业科技股份有限公司(证券代码:839145) 电话:010-62176775
网址:www.hyxyyq.com

Advertisement

Table of Contents
loading

Summary of Contents for Chauvin Arnoux D38N

  • Page 1 GB - User’s manual D38N 北京海洋兴业科技股份有限公司(证券代码:839145) 电话:010-62176775 网址:www.hyxyyq.com...
  • Page 2 ENGLISH Thank you for purchasing a D38N current clamp. For best results from your instrument: „ read this manual carefully, „ comply with the precautions for use. WARNING, DANGER! The operator should refer to this user’s manual whenever this danger symbol appears.
  • Page 3: Table Of Contents

    4. MAINTENANCE ........21 3. SPECIFICATIONS ........17 4.1. Cleaning ........21 3.1. Conditions of reference ....17 5. WARRANTY ..........21 3.2. D38N clamp ........18 3.3. Sine peak current in permanent operation as a function of the frequency .........19 3.4. Environmental conditions ....19 3.5. Construction specifications ....20 北京海洋兴业科技股份有限公司(证券代码:839145)...
  • Page 4: Presentation

    5 busbars of 125 mm x 5 mm or 3 busbars of 100 mm x 10 mm, with the spacing between the busbars equal to their thickness. The D38N clamp is intended for measuring alternating currents and displaying them on an oscilloscope. Clamp...
  • Page 5: Front View

    1.3. FRONT VIEW Jaws. Centring references. Guards. Slider switch. Arms. Parts gripped. Cable. BNC socket. 北京海洋兴业科技股份有限公司(证券代码:839145) 电话:010-62176775 网址:www.hyxyyq.com...
  • Page 6: Operation

    As far as possible, keep away from other conductors which could generate spurious fields. „ Read the measurement on the instrument and apply the reading coefficient to the measurement in V Clamp Coefficient D38N 30 A range 10 mV / A...
  • Page 7: Specifications

    3. SPECIFICATIONS 3.1. CONDITIONS OF REFERENCE Influencing quantity Reference value Temperature 23 ± 3 °C Relative humidity 0 to 85 % RH Frequency of signal measured 48 to 65 Hz Type of signal sinusoidal without DC component Adjacent current‑carrying conductor none Conductor position in jaws centred...
  • Page 8: D38N Clamp

    北京海洋兴业科技股份有限公司(证券代码:839145) 电话:010-62176775 网址:www.hyxyyq.com...
  • Page 9: Sine Peak Current In Permanent Operation As A Function Of The Frequency

    3.3. SINE PEAK CURRENT IN PERMANENT OPERATION AS A FUNCTION OF THE FREQUENCY Ip (A) 10 000 1 000 f (Hz) 1 000 10 000 100 000 3.4. ENVIRONMENTAL CONDITIONS The instrument must be used in the following conditions: % RH 1: Reference range 2: Operating range 3: Storage range...
  • Page 10: Construction Specifications

    3.5. CONSTRUCTION SPECIFICATIONS Dimensions: 310 x 120 x 48 mm Opening: approx. 90 mm Ø 4 mm cable 2 m long terminated by a BNC socket. Weight: approx. 1,200 g. Protection rating: IP 20 as per IEC 60529 3.9. COMPLIANCE WITH INTERNATIONAL STANDARDS The instrument complies with the IEC 61010‑2‑032 standard for type‑A sensors.
  • Page 11: Maintenance

    4. MAINTENANCE The instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may seriously impair safety. 4.1. CLEANING Disconnect the instrument completely. Use a soft cloth, dampened with soapy water.

Table of Contents