Configuration

Before using the Micro Diary the required sequence of
questions and spirometry tests must be defined using the
PC software (Please see Help files on software for
guidance). Micro Diary PC Software may be downloaded
using the link:
http://www.micromedical.co.uk/update/download.asp
(The PC Software is intended for use by Clinicians
and Trial co-ordinators only).
Once defined, the sequence can be loaded into the Micro
Diary. This can be done with a direct connection to a USB
port of the PC: plug the unit into a USB port on the PC,
using the cable provided, and turn the unit on with the
'enter' key depressed to display the operations menu.
Once the operations menu is displayed, release the 'enter'
key. Scroll to option 1, PC Connection, using the up and
down keys and then press 'enter' (the other options are
described in section Menu Options). From the PC
software, download the program to the unit and switch off
by pressing 'del'.
The Micro Diary is now ready to perform the defined
sequence of questions when the unit is turned on. The
subject is instructed to complete the sequence of
questions and spirometry tests. Once complete, the unit
will turn off automatically.
If the subject turns the unit on but leaves the unit before
completing the sequence, the unit will automatically turn