5. Important information
Power connection – Equipment with external 19 VDC power supply
Power requirements: The equipment must be powered using the delivered medical approved 19 VDC
(
) power supply. Model: DELTA MDS-090AAS19 B.
The medical approved DC (
1.5-0.75A 50-60Hz).
The power supply is specified as a part of the ME equipment or combination is specified as a ME
system.
Make sure the voltage of the power source is correct before connecting the equipment to the power
outlet.
The equipment should be installed near an easily accessible outlet.
The equipment is intended for continuous operation.
Transient over-voltage
If the equipment is not used for a long time, disconnect it from the AC inlet to avoid damage by transient
over-voltage.
To fully disengage the power from the equipment, please disconnect the power cord from the AC inlet.

Power cords

Do not overload wall outlets and extension cords as this may result in fire or electric shock.
Power cords should be routed so that they are not likely to be walked upon or pinched by items placed
upon or against them, paying particular attention to cords at plugs and receptacles.
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Grounding
Optionally earth the equipment by connecting the functional earth blind fastener to a grounded outlet by
means of a wire with at minimum AWG18 size (according to national Regulation requirements regarding
maximum admitted cable length).
Water and moisture
Keep this equipment away from excessive humidity.
Never use the equipment near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink,
laundry tub or in a wet basement.
Cleaning
The equipment can be cleaned regularly according to usage. In environments where specific chemi-
cals or cleaning agents are used, please contact your local Barco office.
Do not apply or spray liquid directly to the equipment as excess liquid may cause damage to internal
electronics which may cause fire or electrical shock. Instead, apply the liquid to a cleaning cloth.
External equipment
The RJ11 connector may not be connected to a telephony or telecom network and is for use only with
the hospital's internal network service call system.
If the equipment is equipped with a handset, the handset can magnetically retain small sharp objects
(such as pins & paperclips) around the mouthcap or earcap region, please check and remove before
use.
Accessory equipment connected to the analog and digital interfaces must be in compliance with the
respective nationally harmonized IEC standards (i.e. IEC 60601 for data processing equipment). If in
doubt, consult the technical services department or your local representative.
18
) power supply must be powered by the AC mains voltage (100-240V~
K5903250 JAO15 / JAO18 10/03/2016