5. Important information
5.4

EMC notice

When the JAO15 / JAO18 is used in combination with external peripherals, including
the JAOHandset, the test levels which are listed further below are not applicable and
the EN55022 (CISPR22) and EN55024 (CISPR24) test levels apply instead.
General information
No specific requirement on the use of external cables or other accessories except power supply.
With the installation of the device, use only the delivered power supply or a spare part provided by the
legal manufacturer. Using another can result in a decrease of the immunity level of the device.
Electromagnetic emissions
The JAO15 / JAO18 is intended for use in the electromagnetic environment specified below. The customer
or the user of the JAO15 / JAO18 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
This JAO15 / JAO18 complies with appropriate medical EMC standards on emissions to, and interference
from surrounding equipment. Operation is subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any interference received, including interfer-
ence that may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following mea-
sures:
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The JAO15 / JAO18 is intended for use in the electromagnetic environment specified below. The customer
or the user of the JAO15 / JAO18 should assure that it is used in such an environment.
22
Compliance
Group 1
Class B
Class D
Complies
Electromagnetic environment –
Guidance
The JAO15 / JAO18 uses RF
energy only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
The JAO15 / JAO18 is suitable
for use in all establishments,
including domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.