CHAPTER 9
CALIBRATION
9.1 CALIBRATION INTERVAL
Periodic re-calibration of the ADC-16 is necessary for accuracy, with the interval of recalibration based
on
type
of service. For example, an environment with frequent large changes of temperature and/or
vibration would call for a 3-month re-calibration interval, while laboratory or office conditions would
call for six months to one year.
9.2 CALIBRATION PROGRAM
Your Distribution Software contains the calibration program ADCZ6CAL.EXE is provided to step you
through the calibration process. To run ADClGCAL.EXE, log to the ADC-16 directory and type
ADC16CAL
.
From the opening menu, select as follows:
Change Base Address
-
To find out how to reset the Base Address Switch. The default setting is
Main Menu
-
To access the calibration routines. These routines consist of calibrating the
1
mA
source
(
Set 1 mA Source
)
and calibrating the A/D converter
(
Cal
AID
1.
There is also a utility
program (select
A/D
Utility
);
this program allows a check of the ADC-16 configuration, and
it
performs A/D measurement, and controls the DI and DO.
ADC-16 lnfo
-
To find out more about the ADC-16.
300h (768 decimal).
9.3 REQUIRED TEST EQUIPMENT
Calibration of the ADC-16 requires the following equipment:
Set 1 mA Source Procedure
STA-EX8 and C-1800 cable
DMM (Keithley Instruments 196 or equivalent)
lK,
0.1%,
1/4 watt (min) resistor (optional)
A/D Calibration Procedure
STA-EX8 and C-1800 cable
DC
Voltage Calibrator, EDC type or equivalent
More test equipment specifications are given in the ADC-16 Info screen.
...
9 - 1