Remember to check and, if necessary, adjust the position of the limit stops if the grips are
exchanged for a different holding accessory.
Limit stops on a MultiTest 1-xt. Stops on other stands all operate in the same way.
Twin-column stands have system limit stops as well as limit stops. System limit stops
should not be moved. In an emergency, to release a trapped sample the system limit can
be moved, but if you do so, you must contact your Mecmesin distributor for servicing
and re-setting. These stops are designed for protection from damage, and moving them
may affect your warranty on the stand.
Lower limit stop and system limit stop on a twin-column stand
10  Mecmesin
Assembly and Installation of –xt System Test Stands
upper and lower
limit stops
The crosshead will stop when
it reaches this position
Lower limit stop
System limit stop