Operation Instructions - 3B Medical Cirrus 5 User Manual

Oxygen concentrator
Table of Contents

Advertisement

6. Intake air filter
7. Storage cabinet
For storage of oxygen cannula
8. LCD display (Liquid crystal display)
a. Displays operating status of the oxygen concentrator, refer to 7. on page 9.
b. When starting the oxygen concentrator, the LCD screen will be backlit. It will return to
screen saver mode (not lighted) in 15 minutes
on backlight.
10. Humidifier
(Accessory Item)
11. Air Outlet (optional)
12. Rating label
13. Power Cord Connector
14. Power Cord Holder
9

Operation instructions

: Allow for at least 1 foot (12 inches) of clearance away from walls, draperies, furniture,
or other obstructions. Do not place the unit in a confined, non-ventilated area.
: When turned off, allow at least 5 minutes before restarting concentrator.
9.1. If used with a humidifier, ensure only distilled water is used and the humidifier is filled
properly and the lid is secure prior to operation of the concentrator (Fig 3.). Refer to
humidifier's manufacturer's guidelines for proper filling, maintenance requirements. Place
the humidifier in the proper location (Fig 4). The humidifier must comply with the general
requirement of the 93/42/EEC European Directive as appropriate.
Cirrus 5 User Manual
Pressing any key during operation will turn
.
10

Hide quick links:

Advertisement

Table of Contents
loading