GREISINGER GLF 100 RW Operating Manual

GREISINGER GLF 100 RW Operating Manual

Conductivity measuring instrument

Advertisement

H66.0.02.6C-08
Conductivity Measuring Instrument
(pure water)
GLF 100 RW
Operating Manual
as of version 1.7
WEEE-Reg.-Nr. DE 93889386
GHM Messtechnik GmbH • Standort Greisinger
Hans-Sachs-Str. 26 • D-93128 Regenstauf
 +49 (0) 9402 / 9383-0  +49 (0) 9402 / 9383-33 
info@greisinger.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GLF 100 RW and is the answer not in the manual?

Questions and answers

Summary of Contents for GREISINGER GLF 100 RW

  • Page 1 Conductivity Measuring Instrument (pure water) GLF 100 RW Operating Manual as of version 1.7 WEEE-Reg.-Nr. DE 93889386 GHM Messtechnik GmbH • Standort Greisinger Hans-Sachs-Str. 26 • D-93128 Regenstauf  +49 (0) 9402 / 9383-0  +49 (0) 9402 / 9383-33  info@greisinger.de...
  • Page 2: Table Of Contents

    H66.0.02.6C-08 Operating Manual GLF 100 RW Page 2 of 8 CONTENTS Intended use ......................... 3 General advice ........................3 Safety instructions ....................... 3 Operating and maintenance: ....................3 Disposal notes ........................4 Display elements ........................4 Pushbuttons ......................... 4 ................4 Min-/max value memory ..................
  • Page 3: Intended Use

    H66.0.02.6C-08 Operating Manual GLF 100 RW Page 3 of 8 1 Intended use Precise measurements of conductivity in pure water applications with conductivities below 100µS/cm. Attention: Tap water is no pure water (several 100/µS/cm)! Applications:  Monitoring of boiler water ...
  • Page 4: Disposal Notes

    H66.0.02.6C-08 Operating Manual GLF 100 RW Page 4 of 8 5 Disposal notes Dispense exhausted batteries at destined gathering places. This device must not be disposed as “residual waste”. To dispose this device, please send it directly to us (adequately stamped).
  • Page 5: Configuration Of The Instrument

    H66.0.02.6C-08 Operating Manual GLF 100 RW Page 5 of 8 8 Configuration of the instrument To configure the instruments according to Your needs proceed like follows:  Switch off the instrument.  Switch on the instrument and press mode-key (...
  • Page 6: Adjusting Of The Instrument

    H66.0.02.6C-08 Operating Manual GLF 100 RW Page 6 of 8 9 Adjusting of the instrument The cell correction may change due to natural ageing or deposits at the measuring cell. In case you have an accurate reference liquid, you may adjust the instrument by changing the cells scale adjustment accordingly.
  • Page 7: 12 General Information About Low Conductivity Measuring

    Attention! Wrong handling of reference solution can make them useless very fast. General: Tap water is no pure water and commonly has a conductivity of several 100 µS/cm. The upper range of the GLF 100 RW is 100µS/cm, therefore it is not suitable to measure tap-water. 12.2 Measuring procedure: Before immersion to the measuring solution, rinse the electrode with deionised water, dry with a paper towel and shake remaining water out.
  • Page 8: Specification

    H66.0.02.6C-08 Operating Manual GLF 100 RW Page 8 of 8 13 Specification Measuring Principle Conductivity measuring via 2 pole stainless steel electrode Meas. Ranges Range Resolution Conductivity 0.000 ... 2.000 µS/cm 001 µS/cm 0.00 ... 20.00 µS/cm 0,01 µS/cm 0.0 ... 100.0 µS/cm .