H66.0.02.6C-08 Operating Manual GLF 100 RW Page 3 of 8 1 Intended use Precise measurements of conductivity in pure water applications with conductivities below 100µS/cm. Attention: Tap water is no pure water (several 100/µS/cm)! Applications: Monitoring of boiler water ...
H66.0.02.6C-08 Operating Manual GLF 100 RW Page 4 of 8 5 Disposal notes Dispense exhausted batteries at destined gathering places. This device must not be disposed as “residual waste”. To dispose this device, please send it directly to us (adequately stamped).
H66.0.02.6C-08 Operating Manual GLF 100 RW Page 5 of 8 8 Configuration of the instrument To configure the instruments according to Your needs proceed like follows: Switch off the instrument. Switch on the instrument and press mode-key (...
H66.0.02.6C-08 Operating Manual GLF 100 RW Page 6 of 8 9 Adjusting of the instrument The cell correction may change due to natural ageing or deposits at the measuring cell. In case you have an accurate reference liquid, you may adjust the instrument by changing the cells scale adjustment accordingly.
Attention! Wrong handling of reference solution can make them useless very fast. General: Tap water is no pure water and commonly has a conductivity of several 100 µS/cm. The upper range of the GLF 100 RW is 100µS/cm, therefore it is not suitable to measure tap-water. 12.2 Measuring procedure: Before immersion to the measuring solution, rinse the electrode with deionised water, dry with a paper towel and shake remaining water out.