Override Runtime Cal Factor; Replace; Delete; Setting A Cal Region - Beckman Coulter Navios Instructions For Use Manual

Flow cytometer
Table of Contents

Advertisement

Using Navios Software
Analysis Menu

Override runtime CAL Factor...

Check this option if you want to replay a batch of listmode files using a different Cal Factor.
Advanced
If your Administrator has assigned you access rights, the
enter and modify batch and count information. Enter the
and
CAL Factor.
ADD
Use the ADD button to enter new lot information (only available on the Advanced screen).

Replace

Use the Replace button to replace the selected information with updated values (only available on
the Advanced screen).

Delete

Use the Delete button to remove a selected lot from the list (only available on the Advanced screen).

Setting a CAL Region

Create a Region around the absolute count calibration particles, the region MUST be named CAL. Set
or select the calibration factor for the batch of beads in use from the
Calibration
NOTE
If the Cal Factor is set to 0 or there is no CAL Region present in the protocol, "Error" appears in the
Cells/uL field. If the number of events in the CAL region equal to or less than 1000, "Invalid" appears in
the Cells/uL field.

Calculate Results

This option calculates the statistics for all files selected on the screen.

Refresh

View > Refresh
Selecting
and FlowPAGEs. This option is useful when Windows fails to refresh the screen correctly leaving
some graphical errors visible.
10-144
See
Add Absolute Calibration Batches
menu item.
, or pressing
Advanced
Product name
in the User Administration section.
(
)
, causes Navios software to refresh the screen, all plots
F5
button is active and you can
,
Expiry Date
,
Lot Number/ID
Analysis >
Absolute Count
773232AH