OVERVIEW

MECHANICAL
INSTALLATION
http://www.raihome.com
Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-4
This section describes SPS 4001B installation.
Before starting to install this equipment, read the "Safety instructions for the wiring and
installation of this apparatus" in Section A: Safety Data. Failure to follow the safety
instructions could result in serious injury or death.
Secure equipment cover and ground leads after installation. Failure to secure covers and
ground leads could result in serious injury or death.
The SPS 4001B is only intended for use in general purpose, ordinary locations
(non-hazardous). Do not install the SPS4001B in areas where flammable or ignitable
substabces are present.
The outline drawing in Figure 2-1 shows mounting centers and clearances of
the SPS 4001B. The unit is designed to mount on a wall, bulkhead, or pipe.
Ensure the unit is installed according to the following specifications.
1. Install the unit no further than 300 ft (91 m) from the Oxymitter 4000 and
no further than 1000 ft (303 m) from the electronics package or any
customer-supplied remote input or relay output in the control room.
2. Locate the unit where the ambient temperature is between -25° and
149°F (-5° and 60°C) or -40° and 149°F (-40° and 60°C) with optional
heater.