Monitor Connector; I Monitor Signal; E Monitor Signal; External Signal In - Gamry Interface 1000 Operator's Manual

Potentiostat/galvanostat/zra
Hide thumbs Also See for Interface 1000:
Table of Contents

Advertisement

A special cable is used to connect the instruments that will be synchronized.
connector on one end and a variable number of Serf connectors.
Gamry sells three standard Master-Serf cables:
985-132
985-133
985-134
All of these cables have the Master end of the cable identified by a large M on the DIN connector. Note that a
cable can be used with some instruments missing, so the eight instrument Master -Serf cable can be used for
one Serf, two Serfs, three Serfs,..., seven Serfs.
We recommend that you always plug the Master and last Serf ends of the cable into a
potentiostat.
disconnected.
The Master-Serf connection scheme sends 24 MHz complex clock signal from the Master to the Serfs using a
daisy-chained low voltage differential signal (LVDS). Data acquisition pulses are imbedded in the clock signal.
DC pins in the DIN connector are used to detect all cable connections and most importantly detect the Master
end of the cable.
We ask that you use Gamry supplied cables for all Master - Serf synchronized systems. The cable uses a
differential pair that requires controlled impedance making user construction of a cable very difficult.

Monitor Connector

The Monitor Connector is located on the Interface 1000 Front Panel. It is an 8-pin round mini-DIN connector.
One possible mating connector is from CUI Inc. part number MD-80. The pin out of the Monitor Connector is
shown in Table C-2.
Pin
Name
1
Unreg +5 V
2
Signal Ground
3
Power Ground
4
Signal Ground
5
E_sig
6
EXT_SIG
7
I_Sig
8
Signal Ground

I Monitor Signal

The I Monitor signal represents the output of the Interface 1000 current measurement circuit. It can be treated
as a 2-wire differential signal with the negative side connected to the Interface 1000 Signal Ground. With the
exception of the filtering described below, it is the raw signal with no offset or gain applied. The effective
bandwidth of the current signal falls as you reach the nA and pA current ranges. IE Stability capacitors further
slow the response.
Appendix C -- I/O Connectors--Monitor Connector
One Master and One Serf
One Master and up to Three Serfs
One Master and up to Seven Serfs
You can leave of the DIN connectors in the middle of the cable
Monitor Connector
Use
An unregulated source of +5 V. 10 Ω output impedance.
A very high quality ground. Signals are relative to this ground
Ground used when active circuits are needed. Power circuits from
Pin 1 to Pin 3.
A very high quality ground. Signals are relative to this ground

E Monitor Signal

External Signal In

I_Monitor Signal
A very high quality ground. Signals are relative to this ground
Note
Table C-2
11 - 2
The cable has a Master

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents