Characterizing the delay volumes: Mixing point to UV detector and Column to UV detector
Dead volumes of
The dead volume of some modules is dependent on pressure and solvent;
Agilent pumps
therefore, the table contains a calculation of the dead volumes for
Water/Acetonitrile and Water/Methanol gradients using the initial pump
pressure (p
Pump
G1311A/C
G7111A
1120 Compact LC
1220 VL Compact LC
G1311B
G7111B
1220 Compact LC
G5611A
G1312A/C
G1312B
G7112B
G4204A
G7104A/C
G4220A/B
G7120A
Dead volumes of
Because of the flow-through design, the injected volume must be added to the
Agilent
module's dead volume.
autosamplers
Purification Solution - Developer's Guide
If used, heat exchange capillary in the thermostat of the column
compartment (TCC)
Column to detector
Inlet capillary of the UV detector cell
Dead volume of the UV detector cell
, bar) at the beginning of the gradient.
0
Volume, mL
Water/ACN (p
, bar)
0
p
/2000 + 0.76
0
p
/2760 + 0.742
0
p
/2860 + 0.725
0
p
/3330 + 0.76
0
No mixer: p
/20000 + 0.205
0
Mixer: p
/4000 + 0.77
0
No mixer: 0.46
V380: 0.85
No mixer: 0.033
V35: 0.075
V100: 0.16
V380: 0.39
Calibration Procedures
Volume, mL
Water/MeOH (p
, bar)
0
p
/4000 + 0.82
0
p
/6060 + 0.787
0
p
/2860 + 0.725
0
0.865
No mixer: 0.225
Mixer: p
/6670 + 0.83
0
No mixer: 0.46
V380: 0.85
No mixer: 0.033
V35: 0.075
V100: 0.16
V380: 0.39
6
53