The WiTouch
Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch
®
Immunity test
IEC 60601 test level
Conducted RF
EC 61000-4-6 (Not applicable, no AC mains connection)
IEC 61000-4-6
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the TensTouch® is used exceeds the applicable RF compliance level above, the TensTouch
performance is observed, additional measures may be necessary, such as re-orienting or relocating the TensTouch.
between portable and mobile RF communications equipment and the WiTouch
The WiTouch
Pro is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
®
The customer or the user of the WiTouch
distance between portable and mobile RF communications equipment (transmitters) and the WiTouch
below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
Electromagnetic environment – guidance
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the WiTouch®-
80 MHz to 2.5 GHz
Pro, including cables, than the recommended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d=1.17√P
80 MHz to 800 MHz
d=2.33√P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, ashould be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended separation distances
Pro can help prevent electromagnetic interference by maintaining a minimum
®
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d=1.17√P
0.12
0.37
1.17
3.7
11.7
®
should be observed to verify normal operation. If abnormal
®
Pro
®
Pro as recommended
®
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d=1.17√P
d=2.33√P
0.12
0.23
0.37
0.74
1.17
2.33
3.7
7.37
11.7
23.3
Pro should assure that it is used in such an environment.