Model 88C Resistivity Meter
Instruction Manual
Manual No. 207980, Revision F
Instrument No. 207960
© 2017 Fann Instrument Company, Inc

Advertisement

Table of Contents
loading

Summary of Contents for fann 88C

  • Page 1 Model 88C Resistivity Meter Instruction Manual Manual No. 207980, Revision F Instrument No. 207960 © 2017 Fann Instrument Company, Inc...
  • Page 2 Printed in USA. The information contained in this document includes concepts, methods, and apparatus which may be covered by U.S. Patents. Fann Instrument Company reserves the right to make improvements in design, construction, and appearance of our products without prior notice.
  • Page 3: Table Of Contents

    Figure 3-1 Resistivity Meter, Model 88C ................8 Figure 4-1 Resistivity Chart for NaCl Solutions .............. 11 List of Tables Table 3-1 Resistivity Meter, Model 88C Specifications ............ 7 Table 3-2 Range of Environmental Conditions ..............7 Table 6-1 Model 88 Resistivity Meter Included Parts ............ 14 Table 6-2 Optional Equipment ..................
  • Page 4: Introduction

    Model 88C Instruction Manual Introduction The Model 88C Resistivity Meter measures the resistivity of fluids, slurries, and semisolids according to API recommended practice. Field and laboratory personnel rely on this instrument to evaluate formation characteristics from electric logs. Resistivity is the ability of a material to resist conduction; conductivity is the reciprocal of resistivity.
  • Page 5 Model 88C Instruction Manual HOT SURFACE! Alerts the operator that there is a hot surface and that there is risk of getting burned if the surface is touched. EXPLOSION RISK! Alerts the operator that there is risk of explosion. 207980...
  • Page 6: Safety

    Model 88C Instruction Manual Safety Safe laboratory practices and procedures should be observed while operating and maintaining the Model 88C. Battery Safety On nine volt batteries, the positive and negative terminals are on the same end, next to each other. If these terminals touch anything metal, such as steel wool, coins, or a paper clip, a short circuit could occur, creating heat, sparks, a fire, or an explosion.
  • Page 7: Features And Specifications

    Model 88C Instruction Manual Features and Specifications The Model 88C can measure resistivity in three ranges – 0.01 to 2, 0.1 to 20, and 1.0 to 200 ohm meters. The ohm-meter reading can be converted into parts per million sodium chloride (or equivalent calcium chloride, potassium chloride or other carbonate or hydroxyl salts) using the resistivity chart (Figure 4-1).
  • Page 8: Figure 3-1 Resistivity Meter, Model 88C

    Model 88C Instruction Manual Sample Cell Figure 3-1 Resistivity Meter, Model 88C 207980 Revision F, October 2017...
  • Page 9: Operation

    Model 88C Instruction Manual Operation The Model 88C Resistivity Meter can measure resistivity of fluid samples and semisolid samples (e.g., filter cakes or mud solids). Procedures for both types of samples are listed here. The resistivity chart or nomograph (Figure 4-1) is used to determine the quantity of...
  • Page 10: Semisolid Samples

    Model 88C Instruction Manual Do not use solvents to clean the cell. 6. Using the nomograph (Figure 4-1), convert the resistivity in ohm-meters to concentration of chlorides. Semisolid Samples This procedure explains how to measure resistivity of semisolid samples, such as filter cakes and mud solids.
  • Page 11: Figure 4-1 Resistivity Chart For Nacl Solutions

    Model 88C Instruction Manual Figure 4-1 Resistivity Chart for NaCl Solutions 207980 Revision F, October 2017...
  • Page 12: Troubleshooting And Maintenance

    Model 88C Instruction Manual Troubleshooting and Maintenance Calibration and regular maintenance procedures are described in this section. If more extensive maintenance or service of the instrument is required, please contact Fann Instrument Company. Calibration This procedure is a simple method for checking the instrument. A thorough calibration requires a calibration kit.
  • Page 13: Cleaning

    Model 88C Instruction Manual A battery eliminator (P/N 101710234) is available as a separate purchase. See the parts list for details. Cleaning Always clean the cell immediately after making a measurement. Do not use solvents to clean the cell. Avoid scratching the surface of the cell.
  • Page 14: Parts List

    BATTERY 9 VOLT 206724 KNOB 3/4 DIA X 5/8 HEIGHT w/ 2 SET SCREWS 207262 SAMPLE CELL 207263 PIN CONTACT MODEL 88C RESISTIVITY METER 207962 OVERLAY 88C RESISTIVITY METER 208571 JACK POWER PANEL MOUNT 208806 CIRCUIT BOARD ASSEMBLY AMPLIFIER 88...
  • Page 15: Warranty And Returns

    Fann or, at Fann's option, to the allowance to Customer of credit for the cost of such items. In no event shall Fann be liable for special, incidental, indirect, consequential or punitive damages.