Reading Rates
READING RATES WITH INTERNAL TRIGGER (readings per second)
RATE
S
M
F
1. Sensed automatically at power-up.
2. In 20 mV, 20 ohm, and 200 mA DC ranges.
AUTORANGING
The 8842A autoranges up to the highest ranges in all funtions, down to the 200 mV range in the VDC
and VAC funtions, and down to the 200 Ω ranges in the ohms funtions. To select the 20 mV dc, 20Ω, or
200 mA dc range, press the respective range button (or send the respective range command, if using
the IEEE-488 option).
AUTOMATIC SETTLING TIME DELAY
Time in milliseconds from single trigger to start of A/D conversion, Autorange off.
FUNCTION
VDC
VAC
MA DC
MA AC
Ohms
1.
Difference between first reading and final value for an in-range step change coincident with trigger.
For slow reading rate. 50 counts for medium rate; 10 counts for fast rate.
50 Hz
2
2.08 (.26)
2
16.7 (1.04)
100
RANGE
S
20 mV
342
200 mV
342
2V-1000V
342
All
551
200 mA
342
2000 mA
342
2000 mA
551
20Ω
395
200Ω
395
2 kΩ
322
20 kΩ
342
200 kΩ
141
2000 kΩ
141
20 MΩ
1020
Introduction and Specifications
POWER LINE FREQUECNCY
60 Hz
2
2.5 (.31)
2
20 (1.25)
100
READING RATE
M
F
342
9
61
9
17
9
551
551
342
9
17
9
551
551
395
17
106
17
17
13
17
13
121
21
101
81
964
723
SPECIFICATIONS
1
400 Hz
2
2.38 (.30)
2
19.0 (1.19)
100
NUMBER OF COUNTS
1
FROM FINAL VALUE
30
5
9
30 (Note 2)
9
5
30 (Note 2)
40
5
5
5
5
10
10
1
1-9