Troubleshooting; Optional Accessories - Sanitron S2400C Installation Operation & Maintenance

Table of Contents

Advertisement

TROUBLESHOOTING

WARNING: Always disconnect power to the water purifier before performing any service or maintenance.
IMPORTANT: This unit is to be serviced ONLY by qualified, and appropriately licensed, personnel.
Problem
Purifier not operating
Water leaking into/from purifier.
Poor purifier performance AND/OR Low UV intensity (As
indicated on optional Guardian

OPTIONAL ACCESSORIES

Optional Accessories
Guardian™ Ultraviolet Monitor - Analog/Digital
Sentry™ Safety Sensor
Steralert™
Audio Alarm
Elapsed Time Indicator Universal Input
Time Delay Mechanism
Wall Mounting Kit
Solenoid Valve - (1-1/2")
Solenoid Valve - (2")
Sure Flo™ Flow Control - PVC (2")
Sure Flo™ Flow Control Valve - Stainless Steel (2")
Sure Flo™ Flow Control Valve - PVC (2" Union)
14
Table 2 - Troubleshooting
TM
Ultraviolet Monitor)
Table 3 - Optional Accessories
Available For:
S2400C, S5,000C - S25,000C
S2400C, S5,000C - S25,000C
S2400C, S5,000C - S25,000C
S2400C, S5,000C - S25,000C
S2400C, S5,000C - S25,000C
S2400C, S5,000C - S25,000C
S2400C
S2400C
S5,000C - S25,000C
S2400C
S2400C
S5,000C - S25,000C
Possible Cause
No electrical power....
Cracked or broken quartz sleeve...
Quartz sleeve sealing o-ring(s) worn,
damaged...
Poor, or loose, connections or fittings
Quartz sleeve fouled...
Sensor Probe, if equipped, lens or
quartz rod fouled
Germicidal lamp output depreciating...
Germicidal lamp not functioning...
Low input voltage...
Change in water quality....
Corrective Action
Verify that the purifier is connected to a live power
source.
Shut down purifier, drain, and replace quartz
sleeve. See "Quartz Sleeve Installation or
Replacement" in the "Maintenance" section
Shut down purifier, drain, and remove static
gland nut, replace sealing o-ring. See "Quartz
Sleeve Installation or Replacement" in the
"Maintenance" section
Tighten suspect connection or fitting; or shut down
purifier, drain, and remove fitting or connection.
Clean threads; reapply thread sealing tape and
reinstall.
Clean quartz sleeve, see "Quartz Sleeve
Cleaning" in the "Maintenance" Section.
Clean lens or Quartz Rod, see "Optional
Ultraviolet Monitor Sensor Probe Cleaning" in
the "Maintenance" section
Replace lamp, as it nears its end of life (EOL).
See "Lamp Installation or Replacement" in the
"Maintenance" section.
Replace lamp. See "Lamp Installation or
Replacement" in the "Maintenance" section
Verify input voltage to purifier.
Have water tested to confirm that it does not
exceed maximum recommended concentration
levels for use with this purifier.
Most options are available for
operation at 120v 60Hz or 220v 50Hz.
Please specify.
Not all options may be available
for all models, consult factory for
availability
 Solenoid requires 2-150 PSI for
satisfactory operation.
 Unless otherwise specified PVC
flow control valves are supplied.
All PVC and Stainless Steel flow
control valves are male NPT.
Consult Factory for other flow
control valves