Declassification and Security Instructions
Revision A
www.tektronix.com
071-1999-00

Advertisement

loading

Summary of Contents for Tektronix TDS1001B

  • Page 1: Digital Storage Oscilloscope

    TDS1000B and TDS2000B Series Digital Storage Oscilloscope Declassification and Security Instructions Revision A www.tektronix.com 071-1999-00...
  • Page 2 Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
  • Page 3: Table Of Contents

    Table of Contents Preface ......................Clear and Sanitize Procedures..................Memory Devices....................Data Export Devices................... Troubleshooting..................... How to Clear or Sanitize a Non-Functional Instrument ........... TDS1000B and TDS2000B Series Declassification and Security...
  • Page 4 Table of Contents TDS1000B and TDS2000B Series Declassification and Security...
  • Page 5: Preface

    TDS2022B TDS2024B Related Documents The TDS1000B and TDS2000B Series Digital Storage Oscilloscope Service Manual, part number 071-1828-XX, is available on the Tektronix Web site at www.tektronix.com/manuals. Terms The following terms may be used in this document: Clear. This removes data on media/memory before reusing it in a secured area.
  • Page 6 Preface Sanitize. This eradicates the data from media/memory so that the data cannot be recovered by other means or technology. This is typically used when the device will be moved (temporarily or permanently) from a secured area to a non-secured area.
  • Page 7: Clear And Sanitize Procedures

    Input minimum size Function Modifiable method Location Process to clear SDRAM, Acquisition memory for Only by TDS1001B, TDS1002B, TDS1012B, Remove power from the 512K X 32 holding and processing processor TDS2002B, instrument for at least waveforms, and system. TDS2012B, TDS2022B 20 seconds.
  • Page 8 Input minimum size Function Modifiable method Location Process to clear FLASH, Holds instrument Only by TDS1001B, TDS1002B, TDS1012B, See the Clear Flash 8M X 8, firmware, current setup, processor TDS2002B, Procedure that follows 4M X 16 saved setups, saved system.
  • Page 9 Clear and Sanitize Procedures 13. Insert a USB flash drive into the USB Flash Drive port on the front of the oscilloscope. 14. Push the front panel UTILITY button. 15. Push the Options option button. 16. Push the Printer Setup option button. 17.
  • Page 10: Data Export Devices

    Clear and Sanitize Procedures Data Export Devices The following table lists the data export devices in the standard instrument and listed options. Table 3: Data Export Devices User Input Type Function Modifiable method Location Process to disable Yes. Files can be deleted or overwritten on the USB host Supports removable USB flash Save.
  • Page 11: Troubleshooting

    USB Flash Drive Remove the USB flash drive and return the instrument to Tektronix for repair. After removal of the USB flash drive, refer to your company’s internal policies regarding handling or disposal of the flash drive.