    Bante920 Benchtop pH/ORP Meter Introduction Thank you for selecting the Bante920 benchtop pH/ORP meter. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. UNPACKING THE METER: Before unpacking, ensure that the current work environment meet following conditions.
  • Page 3 Bante920 Benchtop pH/ORP Meter Display Bante920 benchtop pH meter is equipped with a clear and bright LCD display that used to show measured values, mode indicators and help message. The following table describes the meaning of each indicator. INDEX: Measurement mode indicator:...
  • Page 4 Bante920 Benchtop pH/ORP Meter Keypad The meter has a succinct membrane keypad, names and symbols describe the each function key controls. MEAS| ● Power the meter ON/OFF. ● Freezes the measured value on the display, press the key again to resume measuring.
  • Page 5: Connecting The Electrode Holder

    Bante920 Benchtop pH/ORP Meter Connectors Bante920 benchtop pH meter provides 4 connectors for connecting the various types of sensors. Listed in the below table are the details of these connectors. pH/ISE COND USB/ INDEX: CONNECTOR FUNCTION pH/ISE For connecting the pH or ORP electrode...
  • Page 6: Connecting The Electrode

    Bante920 Benchtop pH/ORP Meter Insert the plastic rod into the irregular round hole and swivel the electrode arm 90 . Electrode holder is now ready to swing into desired position. ADJUSTMENT OF ELECTRODE ARM: After installation, if the electrode arm automatically rises or falls, you need to adjust the screws until arm locate at any position.
  • Page 7: Prior To Use

    Bante920 Benchtop pH/ORP Meter Preparing the pH Buffer Solutions Bante920 benchtop pH meter is packaged with three pH buffer packets required for calibration. ● Open the pH7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until reagent is completely dissolved.
  • Page 8: Setup Menu

    Setup Menu Bante920 benchtop pH meter contains an integrated setup menu that allows you to customize each displayed option to meet measurement requirements. In the different modes, the meter will show the corresponding options. For the general options, once it is changed that will be applied to all modes.
  • Page 9: Setting The Default Parameters

    Bante920 Benchtop pH/ORP Meter Auto-Hold: Enable When the auto-hold function is enabled, the meter will automatically sense a stable end-point reading ● Disable and freeze it. 10 minutes Auto-Power Off: 20 minutes When the auto-off power is enabled, if you do not...
  • Page 10: Exit The Setup Menu

    Bante920 Benchtop pH/ORP Meter …………Parameter …………Page number Press  or  key to scroll through menu, select the parameter you want to set (Refer to Setup Menu section). Press ENTER key to confirm, the display shows an option in the submenu.
  • Page 11: Setting The Date And Time

    Bante920 Benchtop pH/ORP Meter Setting the Date and Time Bante920 benchtop pH meter has a real time clock that is used to time-stamp stored measured value and calibration data. Follow the steps below to set the date and time during the first use.
  • Page 12: Automatic Temperature Compensation

    Bante920 Benchtop pH/ORP Meter Press  or  key to set the date and time. Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed. Temperature Compensation In order to get accurate measuring results, you need to enable the manual or automatic temperature compensation before measurement or calibration.
  • Page 13: Single Point Calibration

    Calibration Bante920 benchtop pH meter allows up to 5 points calibration in the pH mode. We recommend that you perform at least 2 points calibration for high accuracy measurement. The meter will automatically recognize and calibrate to following standard buffer values.
  • Page 14: Multi-Point Calibration

    Bante920 Benchtop pH/ORP Meter MULTI-POINT CALIBRATION: Make sure that you have selected 2 to 5 points calibration in the setup menu. Repeat steps 1.2 to 1.5 above. When the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with second point calibration.
  • Page 15 Bante920 Benchtop pH/ORP Meter pH CALIBRATION WITH THE USER-DEFINED BUFFERS: Make sure that you have selected “USER” option in the setup menu. Custom buffer solution values should be at least 1 pH unit apart from each other. Rinse the pH electrode with distilled water. Dip the electrode into the custom buffer solution. Stir the sensor gently. The meter shows the current measured value.
  • Page 16 Bante920 Benchtop pH/ORP Meter pH Calibration Report This program lets you check the parameters of pH electrode for diagnostic purposes. Press MR key in the pH measurement mode, the display shows “LOC/P-01”. Press  or  key until the meter shows “ELE/P-02” (Electrode Diagnosis).
  • Page 17 Bante920 Benchtop pH/ORP Meter Press  key again, the meter shows the slope of each pH buffer group. …..Calibration Points …..Slope After the browsing, press MEAS key to exit the currect mode.  If you are not calibrate the meter or calibration is not successfully, press MR key, the display will show "----" only.
  • Page 18: Orp Calibration

    Bante920 Benchtop pH/ORP Meter ORP Calibration Bante920 benchtop pH meter allows 1 point calibration in the relative mV mode, but calibration is not necessary unless exact readout agreement with a work standard and at a specific ORP value is needed.
  • Page 19 Bante920 Benchtop pH/ORP Meter ORP Calibration Report This program lets you check the millivolt offset in the ORP mode. Press MR key in the ORP measurement mode, the display shows “LOC/P-01”. Press  or  key until the meter shows “ELE/P-02” (Electrode Diagnosis).
  • Page 20: Temperature Calibration

    Bante920 Benchtop pH/ORP Meter Temperature Calibration During the measurement, when automatic temperature compensation is enabled, if the temperature reading displayed differs from that of an accurate thermometer, you need to calibrate the meter. Press and hold the C key for 3 seconds to enter temperature calibration mode, the display shows current temperature reading.
  • Page 21: Orp Measurement

    Dip the electrode into the sample solution, stir the sensor gently. Wait for the reading to stabilize, record the measured value on the display. ORP Measurement Bante920 benchtop pH meter contains two millivolt measurement modes. ABSOLUTE MILLIVOLT MODE: ● Press MODE key until the display shows measurement unit “mV”, the meter is now enters absolute millivolt measurement mode.
  • Page 22: Storing And Recalling Data From Memory

    Bante920 Benchtop pH/ORP Meter Storing and Recalling Data from Memory Bante920 benchtop pH meter allows up to 500 data sets to be stored and recalled. MEMORY INPUT: During the measurement process, press MI key to input measured value into the memory, “Memory” indicator appears on the display.
  • Page 23 Bante920 Benchtop pH/ORP Meter Press ENTER key to confirm, the meter shows page number of the stored data. ..…..…...Page number Press  key, the meter shows date and time of the stored data (Format: mm-dd, hh-mm). Press  key again, the display shows the stored data.
  • Page 24: Receiving Data

    Communication Bante Instruments provides a free Data Acquisition System that can be used to transfer data, receive the measuring values or import the data to Excel. Before using, ensure that Windows 7/8/10 operating system has been installed on your computer.
  • Page 25: Ph Electrode Care And Maintenance

    Bante920 Benchtop pH/ORP Meter pH Electrode Care and Maintenance Since pH electrode is susceptible to dirt and contamination, clean as necessary depending on the extent and condition of use. AFTER MEASURING: Rinse the pH electrode in distilled water, store the electrode into the electrode storage solution.
  • Page 26: Sample Type

    Bante920 Benchtop pH/ORP Meter Optional pH Electrodes Bante920 benchtop pH meter is equipped with a general purpose pH electrode, if it can not fully meet your measurement requirements, please refer to the table below to select an applicable sensor. SAMPLE TYPE ●...
  • Page 27: Specifications

    Bante920 Benchtop pH/ORP Meter Specifications Model Bante920 Range -2.000~20.000pH Accuracy ± 0.002pH Resolution 0.1, 0.01, 0.001pH Calibration Points 1 to 5 points Calibration Solutions USA, NIST, DIN or User-Defined Buffer Temperature Compensation 0~100° C, 32~212° F, Manual or Automatic Range -1999.9~1999.9mV...
  • Page 28 Hazardous Substance Statement Bante Instruments Limited is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like.