Bante900P Portable Multiparameter Water Quality Meter Introduction Thank you for selecting the Bante900P portable multiparameter water quality meter. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Unpacking the Meter Before unpacking, ensure that the current work environment meet following conditions.
Page 3
Bante900P Portable Multiparameter Water Quality Meter Display Bante900P portable multiparameter water quality meter is equipped with a clear and bright LCD display that used to show measured values, mode indicators and help message. The following table describes the meaning of each indicator. INDEX: Measurement mode indicator: Low Battery Alarm:...
Page 4
Bante900P Portable Multiparameter Water Quality Meter Keypad The meter has a succinct membrane keypad, names and symbols describe the each function key controls. INDEX: DESCRIPTION ● Power the meter ON/OFF. MEAS| ● Freezes the measured value on the display, press the key again to resume measuring. ●...
Bante900P Portable Multiparameter Water Quality Meter Connectors Bante900P portable meter provides 3 connectors for connecting the various types of sensors. Listed in the below table are the details of these connectors. INDEX: CONNECTOR FUNCTION BNC Connector For connecting the pH, ORP or Ion Selective Electrodes Phone Jack For connecting the temperature probe 6-pin Connector...
Bante900P Portable Multiparameter Water Quality Meter Connecting the Sensor Take out the sensor from the carrying case. FOR THE pH, ORP OR ION SELECTIVE ELECTRODES: ● Insert the BNC connector into corresponding connector socket. Rotate and push the connector clockwise until it locks. FOR THE CONDUCTIVITY ELECTRODE OR DISSOLVED OXYGEN PROBE: ●...
Page 7
Bante900P Portable Multiparameter Water Quality Meter Preparing the pH Buffer Solutions Bante900P portable multiparameter water quality meter is packaged with three pH buffer packets required for calibration. ● Open the pH7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until reagent is completely dissolved.
Bante900P Portable Multiparameter Water Quality Meter Prior to Use Remove the protective cap from the bottom of the sensor. ● FOR THE pH ELECTRODE: If the glass sensitive membrane on the electrode dries out, soak the electrode in 3M KCL solution or tap water for at least 15 minutes. DO NOT use distilled or deionized water, it will shorten the life of sensor.
Bante900P Portable Multiparameter Water Quality Meter Setup Menu Bante900P portable multiparameter water quality meter contains an integrated setup menu that allows you to customize each displayed option to meet measurement requirements. In the different modes, the meter will show the corresponding options. For the general options, once it is changed that will be applied to all modes.
Page 10
Bante900P Portable Multiparameter Water Quality Meter ● 2 points 3 points Calibration Points: Select the number of calibration points. 4 points 5 points ● Monovalent Ion Valence: Select the ion valence of sensor. Divalent CONDUCTIVITY/TDS/SALINITY/RESISTIVITY MODES: PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT K=0.1 ●...
Page 11
Bante900P Portable Multiparameter Water Quality Meter DISSOLVED OXYGEN MODE: PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT ● 1 point Calibration Points: Select the number of calibration points. 2 points Pressure Coefficient: Setting Range: 450 to 850mmHg Select the default atmospheric pressure coefficient. Salinity Coefficient: Setting Range: 0.0 to 50.0ppt Select the salinity coefficient of sample solution.
Bante900P Portable Multiparameter Water Quality Meter Calibration Due: 1 to 31 days … When calibration due reminder is enabled, if you do not recalibrate meter within a specified time period, ● Disable the meter will automatically show indicator. Date and Time: Sets the date and time of the meter.
Bante900P Portable Multiparameter Water Quality Meter 3. Press or key to scroll through menu, select the parameter you want to set (Refer to Setup Menu section). 4. Press ENTER key to confirm, the display shows an option in the submenu. …………Option Enter …………Parameter...
Bante900P Portable Multiparameter Water Quality Meter Setting the Date and Time Bante900P portable meter has a real time clock that is used to time-stamp stored measured value and calibration data. Follow the steps below to set the date and time during the first use. 1.
Bante900P Portable Multiparameter Water Quality Meter 6. Press or key to set the date and time. 7. Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed. Temperature Compensation In order to get accurate measuring results, you need to enable the manual or automatic temperature compensation before measurement or calibration. AUTOMATIC TEMPERATURE COMPENSATION: ●...
Bante900P Portable Multiparameter Water Quality Meter pH Calibration Bante900P portable multiparameter water quality meter allows up to 5 points calibration in the pH mode. We recommend that you perform at least 2 points calibration for high accuracy measurement. The meter will automatically recognize and calibrate to following standard buffer values. USA Standard Buffer Options pH1.68, 4.01, 7.00, 10.01, 12.45 NIST Standard Buffer Options...
Bante900P Portable Multiparameter Water Quality Meter MULTI-POINT CALIBRATION: Make sure that you have selected 2 to 5 points calibration in the setup menu. Repeat steps 1.2 to 1.5 above. When the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with second point calibration.
Page 18
Bante900P Portable Multiparameter Water Quality Meter pH CALIBRATION WITH THE USER-DEFINED BUFFERS: Make sure that you have selected “USER” option in the setup menu. Custom buffer solution values should be at least 1 pH unit apart from each other. Rinse the pH electrode with distilled water. Dip the electrode into the custom buffer solution. Stir the sensor gently. The meter shows the current measured value.
Page 19
Bante900P Portable Multiparameter Water Quality Meter pH Calibration Report This program lets you check the parameters of pH electrode for diagnostic purposes. 1. Press MR key in the pH measurement mode, the display shows “LOC/P-01”. 2. Press or key until the meter shows “ELE/P-02” (Electrode Diagnosis). 3.
Page 20
Bante900P Portable Multiparameter Water Quality Meter 5. Press key again, the meter shows the slope of each pH buffer group. …..Calibration Points …..Slope 6. After the browsing, press MEAS key to exit the currect mode. If you are not calibrate the meter or calibration is not successfully, press MR key, the display will show "----" only.
Bante900P Portable Multiparameter Water Quality Meter ORP Calibration Bante900P portable multiparameter water quality meter allows 1 point calibration in the relative mV mode, but calibration is not necessary unless exact readout agreement with a work standard and at a specific ORP value is needed. 1.
Page 22
Bante900P Portable Multiparameter Water Quality Meter ORP Calibration Report This program lets you check the millivolt offset in the ORP mode. 1. Press MR key in the ORP measurement mode, the display shows “LOC/P-01”. 2. Press or key until the meter shows “ELE/P-02” (Electrode Diagnosis). 3.
Bante900P Portable Multiparameter Water Quality Meter Ion Concentration Calibration Bante900P portable multiparameter water quality meter supports ion concentration calibration up to 5 points with minimum of 2 points, available calibration points include the following options. MEASUREMENT UNITS CALIBRATION POINTS 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000 mg/L 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000 mol/L...
Page 24
Bante900P Portable Multiparameter Water Quality Meter 3. Press CAL key, the meter shows 0.001ppm (or mg/L, mol/L, mmol/L). 4. If necessary, press or key to select the desired calibration point (e.g., 0.01ppm). 5. Rinse the Ion Selective Electrode with distilled water, then rinse with a small amount of ion standard solution. 6.
Page 25
Bante900P Portable Multiparameter Water Quality Meter 8. Rinse the Ion Selective Electrode with distilled water. Dip the electrode into corresponding calibration solution. Stir the sensor gently. 9. Press ENTER key, “Calibration” indicator begins flashing. Wait for the measured value to stabilize, the display shows “1/CAL3”. The meter prompts you to continue with third point calibration.
Page 26
Bante900P Portable Multiparameter Water Quality Meter Ion Calibration Report This program lets you check the slope of the Ion Selective Electrode. 1. Press MR key in the ion measurement mode, the meter shows “LOC/P-01”. 2. Press or key until the display shows “ELE/P-02” (Electrode Diagnosis). 3.
Bante900P Portable Multiparameter Water Quality Meter Conductivity Calibration Bante900P portable multiparameter water quality meter is capable of using the three types of conductivity electrodes. Before calibration or measurement, you need to select the different sensors as per your sample concentration. The table below shows available sensors and its effective measuring ranges. CONDUCTIVITY ELECTRODE MEASURING RANGES CELL CONSTANT...
Page 28
Bante900P Portable Multiparameter Water Quality Meter Dip the conductivity electrode into the calibration solution, the meter automatically shows current calibration standard (e.g., 1413µS/cm). Press ENTER key to confirm, “Calibration” indicator begins flashing. Enter Wait for the measured value to stabilize, the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed.
Bante900P Portable Multiparameter Water Quality Meter MULTI-POINT CALIBRATION: Make sure that you have selected 2 to 5 points calibration in setup menu. Repeat steps 1.2 to 1.5 above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with second point calibration.
Page 30
Bante900P Portable Multiparameter Water Quality Meter Press and hold the key for 3 seconds to enter setup menu, the display shows “CELL” option. Press ENTER key to confirm, the meter enters the cell constant setting mode. Enter Press or key to select “USER” option. ...
Page 31
Bante900P Portable Multiparameter Water Quality Meter Press ENTER key to confirm, the meter enters User-Defined mode. Enter Press or key to set each digit according to previous record. Press ENTER key to confirm until the setting values stop flashing. The meter returns to measurement mode automatically. Setting is completed. ...
Page 32
Bante900P Portable Multiparameter Water Quality Meter Conductivity Calibration Report This program lets you check the calibration factor of the conductivity electrode. 1. Press MR key in the conductivity measurement mode, the display shows “LOC/P-01”. 2. Press or key until the meter shows “ELE/P-02” (Electrode Diagnosis). 3.
Page 33
Bante900P Portable Multiparameter Water Quality Meter DO Calibration in % Saturation Mode Bante900P portable multiparameter water quality meter can be calibrated quickly in air. In the percentage saturation mode, the meter is able to perform either 1 or 2 points calibration. For single point calibration, we recommend that you perform 100% saturation calibration in the air. For the 2 points calibration, you need to use saturated anhydrous sodium sulfite solution (zero oxygen solution).
Page 34
Bante900P Portable Multiparameter Water Quality Meter 2 POINTS CALIBRATION: Make sure that you have selected 2 points calibration in the setup menu. Press CAL key, the meter shows “100/CAL1”. Press or key until the display shows “0/CAL1”. Dip the dissolved oxygen probe into the saturated anhydrous sodium sulfite solution (zero oxygen solution).
Page 35
Bante900P Portable Multiparameter Water Quality Meter Dip the dissolved oxygen probe into air-saturated water for 3 to 5 minutes, press ENTER key to confirm. Wait for the measured value to stabilize, the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed. Enter ...
Page 36
Bante900P Portable Multiparameter Water Quality Meter Dip the dissolved oxygen probe into the air-saturated water for 3 to 5 minutes. Press ENTER key to confirm, “Calibration” indicator begins flashing. Enter Wait for the measured value to stabilize, the display will show “END”. The meter returns to measurement mode automatically. Single point calibration is completed.
Page 37
Bante900P Portable Multiparameter Water Quality Meter Dip the dissolved oxygen probe into the saturated anhydrous sodium sulfite solution (zero oxygen solution). Stir the probe gently. Press ENTER key to confirm, “Calibration” indicator begins flashing. Enter Wait for the measured value to stabilize, the display shows “8.25/CAL2”. The meter prompts you to continue with second point calibration. Dip the dissolved oxygen probe into air-saturated water for 3 to 5 minutes, press ENTER key to confirm.
Bante900P Portable Multiparameter Water Quality Meter Temperature Calibration During the measurement, when automatic temperature compensation is enabled, if the temperature reading displayed differs from that of an accurate thermometer, you need to calibrate the meter. 1. Press and hold the C key for 3 seconds to enter temperature calibration mode, the display shows current temperature reading.
Bante900P Portable Multiparameter Water Quality Meter pH Measurement 1. Press MODE key until the meter shows indicator. 2. Rinse the pH electrode with distilled water to remove any impurities adhering to the probe body. 3. Dip the electrode into the sample solution, stir the sensor gently. 4.
Bante900P Portable Multiparameter Water Quality Meter Ion Concentration Measurement 1. Press MODE key until the meter shows indicator. 2. Rinse the Ion Selective Electrode thoroughly with distilled water. Dip the electrode into the sample solution. 3. If your sample is belong to low concentration liquids or some interfering ions are present in the solution, we suggest you that adding the Ionic Strength Adjuster into the sample solution.
Bante900P Portable Multiparameter Water Quality Meter Dissolved Oxygen Measurement Bante900P portable multiparameter water quality meter is suitable for measuring the water, wastewater, brine and other liquids. If you use meter to measure seawater or other water containing large amounts of salt, please set the salinity coefficient before use. Some gas and steam such as chloride, sulfur dioxide, sulfureted hydrogen, ammonium, carbon dioxide and iodin can permeate the membrane via diffusion.
Bante900P Portable Multiparameter Water Quality Meter Storing and Recalling Data from Memory Bante900P portable multiparameter water quality meter allows up to 500 data sets to be stored and recalled. MEMORY INPUT: During the measurement process, press MI key to input measured value into the memory, “Memory” indicator appears on the display. ...
Page 43
Bante900P Portable Multiparameter Water Quality Meter 3. Press key, the meter shows date and time of the stored data (Format: mm-dd, hh-mm). 4. Press key again, the display shows the stored data. 5. After the browsing, press MEAS key to exit the current mode.
Communication Software Bante Instruments provides a free Data Acquisition System that can be used to transfer stored data, receive the measuring values or set the interval time for automatic recording. The system requirement is Windows XP or Windows 7. You can download this software from our official website: www.bante-china.com...
Bante900P Portable Multiparameter Water Quality Meter PRINT THE REPORT: If you would like to print current data, click the Print button, the printer will output data sheet. DATA RECALL: DAS data acquisition system has a large database used to store the all uploaded data. If you need to view previous data, click the Recall button, the screen will show all data in the database.
Page 46
Bante900P Portable Multiparameter Water Quality Meter CREATE THE EXCEL DATA FILE: When the transfer is complete, click the “Import data into Excel” button, the measured value in the data sheet will be converted to Excel file. You can follow the path below to find the Excel file: C:>Program Files>Data Acquisition System>Excel NOTE: ●...
Bante900P Portable Multiparameter Water Quality Meter pH Electrode Care and Maintenance Since pH electrode is susceptible to dirt and contamination, clean as necessary depending on the extent and condition of use. AFTER MEASURING: Rinse the pH electrode in distilled water, store the electrode into the electrode storage solution. CLEANING THE ELECTRODE: ●...
Page 48
Bante900P Portable Multiparameter Water Quality Meter Conductivity Electrode Care and Maintenance ● Ensure that the conductivity electrode is thoroughly washed with distilled water after each use. ● If there is a build-up of solids inside the measurement area of the cell, these should be removed very carefully with a cotton bud soaked in solvent, taking care not to touch the metal parts of the inner cell.
Bante900P Portable Multiparameter Water Quality Meter Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION Soak the pH electrode in 3M KCL solution for 15 minutes Electrode dried out Soak the Ion Selective Electrode in the diluted standard solution Soak the conductivity electrode in tap water for a few minutes Measured value is out of range Check the electrode whether clogged, dirty or broken Incorrect calibration solutions...
Page 50
Bante900P Portable Multiparameter Water Quality Meter Range 0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm Accuracy ± 0.5% F.S Calibration Points 1 to 5 points Calibration Solutions 10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm Conductivity Temperature Compensation 0~100° C, 32~212° F, Manual or Automatic Temperature Coefficient 0.0~10.0%/° C Compensation Modes Linear or Pure Water Cell Constant...
Page 51
Bante900P Portable Multiparameter Water Quality Meter Hold Function Manual or Automatic Stability Conditions Low or High Calibration Due 0 to 31 days Power Off Manual or Automatic (10, 20, 30 minutes) Reset Function Memory Stores up to 500 data sets Others Output USB Communication Interface...
Bante900P Portable Multiparameter Water Quality Meter Addendum 1: Optional pH Electrodes Bante900P portable multiparameter water quality meter is equipped with a general purpose pH electrode, if it can not fully meet your measurement requirements, please refer to the table below to select an applicable sensor. SAMPLE TYPE ●...
Page 53
Bante900P Portable Multiparameter Water Quality Meter Addendum 2: Optional Ion Selective Electrodes The meter is capable of connecting a variety of composite ion selective electrodes. The following list shows the selectable sensors and its range. MODEL RANGE F-US Fluoride 1× 10 M~ Saturation CL-US Chloride...
Page 54
Bante900P Portable Multiparameter Water Quality Meter Addendum 5: How to calculate the temperature coefficient To determine the temperature coefficient of sample solution use this formula: × 100% -25)-C -25) Where: =Temperature coefficient =Conductivity at Temperature A =Conductivity at Temperature B =Temperature A =Temperature B 1.
Hazardous Substance Statement Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like.