1. Press <Tests> then highlight and <PICK> [On-ear measures], then highlight and
<PICK> [RECD].
2. If no coupler response exists or the measurement interval has been exceeded, go to 15.3:
RECD coupler responseIf a valid coupler response exists, highlight and <PICK>
[Measure real ear].
3. Refer to Positioning the probe tube and insert the probe tube into the ear canal to within
2 - 5 mm of the eardrum.
4. Insert the foam tip into the ear, being careful not to advance the probe tube further into
the ear canal. Insertion depth should be as used with insert phones. Allow the foam tip
to fully expand in the ear.
5. Press <Continue> to generate the real-ear response curve (pink) and the RECD curve