microphone may be analyzed by 1/12th octave, A-weighted or C-weighted filters. A and C-
weighted measurements are limited to frequencies above 100 Hz and the averaging time
and update rate are 384 ms.
1. Press <Tests>, then highlight and <PICK> [Test box measures], then highlight and
<PICK>[Manual control].
2. Highlight and <PICK> the Level window, then highlight and <PICK> 0 dB stimulus level
from the drop-down list.
3. Highlight and <PICK> the Filter window, then highlight and <PICK> 1/12th octave, A-
weighted, or C-weighted from the drop-down list.
4. For 1/12th octave filter, highlight and <PICK> the Frequency window, then highlight
and <PICK> the desired frequency from the drop-down list.
5. Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To
stop the measurement, highlight and <PICK> [Stop test]. You may change Level and
Frequency without stopping the test.
User's Guide Version 3.4 © June 2011
®
48