Summary of Contents for Covidien Genius 2

  • Page 1 Operating Manual Table of Contents Genius Tympanic Thermometer and Base...
  • Page 2 Operating Manual Genius Tympanic Thermometer and Base...
  • Page 3 Conformity Declaration This product contains software solely owned by Covidien llc. Covidien llc grants the user a non-exclusive, limited license to use the software according to the operating instructions. A copy of the license can be obtained from Covidien llc.
  • Page 4: Section I Introduction

    • If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the respective system. • When first used, the Genius 2 Tympanic Thermometer and Base will show the factory preset settings: ear mode (EAR) and the celsius scale (°C).
  • Page 5: Section Iv Icon Identification

    • Always install a new probe cover prior to taking a temperature. The probe cover membrane should be smooth with no holes, tears, or wrinkles. • The thermometer was designed to be used only with Genius 2 Tympanic Thermometer and Base probe covers. Use of probe covers manufactured by any company other than Covidien llc may result in erroneous readings.
  • Page 6: Section V Instructions For Use

    Peak Select System The Genius 2 Tympanic Thermometer and Base uses the patented Peak Select System. This system is well known from our Genius Thermometer. It reduces technique errors by taking multiple readings and automatically selecting the most accurate temperature for display.
  • Page 7: Probe Covers

    The Genius 2 Tympanic Thermometer and Base utilizes a single use disposable probe cover during temperature measurement. The probe cover was designed especially for use with the Genius 2 Tympanic Thermometer and Base. The use of probe covers produced by any company except Covidien llc will result in erroneous temperature measurement. In order to aid in infection prevention and control, always install a new probe cover prior to taking a temperature.
  • Page 8 English 4. Inspect the probe cover to make sure it is fully seated (no space between cover and tip base) and no holes, tears, or wrinkles are present in the plastic film. 5. Place the probe in the ear canal and seal the opening with the probe tip. For consistent results, ensure that the probe shaft is aligned with the ear canal.
  • Page 9: Temperature Recall

    Sleep Mode The Genius 2 Tympanic Thermometer and Base enters sleep mode after 30-40 seconds of non-use. To wake up the thermometer, eject the attached probe cover and/or install a new probe cover. This energy conserving function optimizes battery life.
  • Page 10: Biotech Mode

    English If the display shows a system error 4 and the system error icon, the ambient conditions are changing too rapidly for the device to be used. Allow the device to equilibrate for 20 minutes before using. If the display shows any other system error, then reset the thermometer by installing a probe cover. If the system error does not clear, then contact the factory for available service options.
  • Page 11: Section Vi Preventative Maintenance

    30 minutes prior to use. • Use of a clean cloth or sponge is recommended for cleaning. Never use an abrasive pad or an abrasive cleaner on the Genius 2 Tympanic Thermometer and Base.
  • Page 12: Section Viii Battery Replacement

    Section VIII — Battery Replacement The Genius 2 Tympanic Thermometer and Base batteries should be replaced when the low battery icon is shown on the LCD display. After the low battery icon is displayed, the user will be able to take approximately 100 temperatures before the dead battery icon is displayed and the unit will not allow a temperature to be taken.
  • Page 13 Clinical Repeatability: Meets Section A.5 of EN 12470-5: 2003 (E) per Covidien llc technical report. Data is available from Covidien llc on request. Response Time: Less than 2 seconds...
  • Page 14: Section Xii Customer Service

    The Genius 2 Tympanic Thermometer and Base follows international regulatory standards. • EN 12470-5:2003 (E) Clinical Thermometers – Part 5: Performance of infrared ear thermometers where sections 6.3.4 and 6.5.5 were met based on Covidien llc technical report. Data is available from Covidien llc on request. • UL 60601-1 •...
  • Page 15: Section Xiii Warranty

    Limited Warranty: Covidien llc warrants to the original purchaser (“Customer”) that this product will be free of defects in materials and workmanship, under normal use, for three (3) years from the date of original purchase from Covidien llc or its authorized distributor.
  • Page 16 Guidance and manufacturer’s declaration – electromagnetic immunity The Genius 2 Tympanic Thermometer and Base is intended for use in the electromagnetic environment specified below. The user of the Genius 2 Tympanic Thermometer and Base should assure that it is used in such an environment.
  • Page 17 Genius 2 Tympanic Thermometer and Base The Genius 2 Tympanic Thermometer and Base is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Genius 2 Tympanic Thermometer and Base can help prevent...
  • Page 18 The Genius 2 Tympanic Thermometer and Base is intended for use in the electromagnetic environment specified below. The customer or the user of the Genius 2 Tympanic Thermometer and Base should assure that it is used in such an environment.
  • Page 19 Определение вещества, не содержащегося или не присутствующего в продукте и упаковке. Ürün veya ambalajında bulunmayan veya var olmayan bir madde tanımlaması. registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. © Made in USA. Tympanic Thermometer and Base...