Section 2

2.1

Services required

The fuel and air supplies to the instrument must be clean and dry and supply pressures
regulated within the limits specified. Any contamination, moisture or variation in supply pressure
will directly affect the performance of the instrument.
NOTE: The instrument will only operate with the correct type of air compressor and gas
regulator.
2.1.1
Voltage
90 - 125V or 190 - 250V @ 50 or 60Hz.
2.1.2
Fuel
Propane or butane regulated to 10-15 inches water gauge (0.36-0.54psi or 0.025-
2
0.038kg/cm
).
Natural gas at mains pressure between 3 and 10 inches water gauge (0.11-0.36psi or
2
0.0076-0.025kg/cm
).
L.P.G. as for propane.
Suitable regulators are available from Jenway; refer to Section 2.5.
2.1.3
Air
A supply of dry, clean and pulse-free air at a pressure between 14 and 30 psi (approx. 1-2
2
kg/cm
) at 6 litres/minute is required. A suitable compressor and water separator are available
from Jenway; refer to Section 2.5.
2.1.4
Drain
The instrument will need to be sited near a drain or sink to enable disposal of waste liquid. A
suitable receptacle can be utilised if provision is made for easy disposal of its contents.
WARNING: The waste liquid will still contain any hazardous materials that were in the original
samples and should be handled and disposed of with the same care. Waste liquid should
always be considered to be of a pathogenic nature where the instrument is used in a clinical
environment.
2.2

Unpacking

Remove the instrument from the packaging and ensure the following items are present:
Model PFP7 or PFP7/C Flame photometer
Auxiliary power plug
Mains cable with connector (plug optional as ordered)
Nebuliser inlet tube (500mm)
Gas tube, low pressure (2 metres)
Silicon rubber tube (drain)
Air tube (2 metres)
Nozzle 1: gas
Nozzle 2: air
Allen key, 2.5mm
Adjustable hose clip
Drain trap
Drain trap clip
Nebuliser cleaning wire
Any shortages or damage should be reported immediately to the manufacturer or local
distributor.

Installation

Part Code
009 035
013 046
500 193
500 191
023 003
500 192
026 012
026 013
060 037
060 083
500 018
500 114
500 194