[Read the Important Safety Rules section before operating this equipment.]
6. Insert power cord into the electrical outlet.
7. Locate the power switch on the front of the unit, and switch it to
the I position (on). (Figure 2.)
A battery-operated audible alarm must sound for a 5-second
test to indicate a good battery and alarm.
If the alarm is weak or does not sound at all, consult
your Equipment Provider immediately.
The standard NewLife Elite Oxygen Concentrator
accommodates prescriptions from 1 to 5 lpm.
8. Dual Flow and High Flow Applications: Set the flowmeter adjustment
knob(s) to the prescribed lpm, in any combination of flows up to a total
of 5 lpm for Dual 5 units (Figure 3).
or
Pediatric/Low Flow Applications: Set the flowmeter adjustable knob(s)
to the prescribed lpm up to 2 lpm in 1/8 liter (125 ccm) increments.
The concentrator is now ready for use.
On/Off
Power Switch
Figure 2
Figure 3
10
MN078-1 Rev F
8/12