Service Manual RSA3408B 8 GHz Real-Time Spectrum Analyzer 071-2366-00 This document applies to firmware version 4.0 and above. Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so.
Page 3
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located.
Page 5
........4- - 33 RSA3408B Analyzer Service Manual...
Page 6
........7- - 3 RSA3408B Analyzer Service Manual...
Page 9
........7- -29 RSA3408B Analyzer Service Manual...
Page 10
......5- -78 Table 5- -8: Symptoms and faulty modules ..... 5- -79 RSA3408B Analyzer Service Manual...
Page 11
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation. RSA3408B Analyzer Service Manual...
Page 12
CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product: Mains Connected Protective Ground CAUTION Mains Disconnected ON (Power) (Earth) Terminal Refer to Manual OFF (Power) viii RSA3408B Analyzer Service Manual...
Page 13
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components. To avoid electric shock, do not touch exposed connections. RSA3408B Analyzer Service Manual...
Page 14
Service Safety Summary RSA3408B Analyzer Service Manual...
Page 15
Preface This is the service manual for the RSA3408B 8 GHz Real-Time Spectrum Analyzer. This manual contains information needed to service the analyzer to the module level. Manual Structure This manual is divided into sections, such as Specifications and Theory of Operation.
Page 16
Related Documentation Other documentation for the RSA3408B analyzer includes: H The RSA3408B User Manual contains a tutorial to quickly describe how to operate the analyzer. It also includes an in-depth discussion on how to more completely use the analyzer features.
Page 17
H Read the General Safety Summary and the Service Safety Summary, beginning on page vii. H Read the RSA3408B User Manual (Tektronix part number 071-2364-xx) for operating information. When using this manual for servicing, be sure to follow all warnings, cautions, and notes.
Page 18
Introduction RSA3408B Analyzer Service Manual...
The specifications for this instrument are available on the Documents CD (Tektronix part number 063-4089-xx) that shipped with your product. Look for the RSA3408B 8 GHz Real-Time Spectrum Analyzer Specifications and Performance Verification Technical Reference PDF (Tektronix part number 071-2480-xx), available on this disk. For the most current documentation, refer to the Tektronix Web site (www.tektronix.com).
Page 22
Specifications 1- 2 RSA3408B Analyzer Service Manual...
Theory of Operation This section describes the electrical operation of the RSA3408B analyzer using the major circuit blocks or modules shown in Figure 6- -1 on page 6- -3. RF Input The RF input circuit block consists of a relay for signal switching (Input Relay), Step Attenuator, Programmable Bandpass Filter, and a control circuit for these components.
Page 26
LO signal and A/D clock signal generated in the Oscillator block are highly stabilized signals locked with a 10 MHz reference signal, and are output to the 3 Converter and A10 A/D board, respectively. 2- 2 RSA3408B Analyzer Service Manual...
In addition, re-sampling is performed to achieve the higher frequency resolution. I/Q signals output from the Digital Filters are sent to the Trigger Detector and Acquisition Memory. 2- 3 RSA3408B Analyzer Service Manual...
The Power Supply 1 output voltages are +5 V / +15 V / - -15 V / +24 V; the Power Supply 2 output voltages are +5 V / +12 V / - -12 V / +3.3 V. 2- 4 RSA3408B Analyzer Service Manual...
CPU board are performed on this board. A62 Front Connection The A62 Front Connection board interfaces with peripheral devices such as hard Board disk drive, floppy disk drive, LCD (Liquid Crystal Display), USB, and keys with the CPU board. 2- 5 RSA3408B Analyzer Service Manual...
Page 30
Theory of Operation 2- 6 RSA3408B Analyzer Service Manual...
Diagnostics just referenced. Advantages: These procedures add direct checking of warranted specifica- tions. Disadvantages: They require more time to perform and suitable test equipment is required. (Refer to Equipment Required on page 3- -10.) 3- 1 RSA3408B Analyzer Service Manual...
The example step below is telling you to set the analyzer controls by pressing the specified key sequence. Set the RSA3408B analyzer controls: Press Mode: Demod > Digital Demod > Constellation. STOP. “STOP” is accompanied by information you must read to do the procedure properly.
The Diagnostics use internal routines to verify that the instrument functions properly and passes the internal circuit tests. The Functional Tests begin on page 3- -4 and the Diagnostics procedures are on page 3- -8. 3- 3 RSA3408B Analyzer Service Manual...
Press Input > Signal Input Port... > Cal100M. The spectrum of the calibration signal appears. d. Check that “INPUT: CAL” and “FREE RUN” are displayed in the status indicator at the upper right of the screen (see Figure 3- -1). 3- 4 RSA3408B Analyzer Service Manual...
Press the Span key on the front panel. b. Confirm that the span is 36 MHz and the RBW is 100 kHz in the setup display on the upper part of the screen (see Figure 3- -2). 3- 5 RSA3408B Analyzer Service Manual...
Figure 3- -3. e. Using the numeric keypad, set the reference level back to 0 dBm. (Press 0 > Enter, in that order, on the keypad.) 3- 6 RSA3408B Analyzer Service Manual...
Press the Run/Stop key on the front panel to stop data acquisition. Confirm that the trace display freezes and PAUSE is displayed in the status indicator at the top right of the screen. 3- 7 RSA3408B Analyzer Service Manual...
You can press the All side key to run all the tests. 5. Check the result shown in the lower left view (“Pass” or “Fail”). The process is graphically displayed on the upper views. Figure 3- 5: Diagnostic screen 3- 8 RSA3408B Analyzer Service Manual...
Performance Tests This section contains procedures for checking that the RSA3408B 8 GHz Real-Time Spectrum Analyzer performs as warranted. The procedures are arranged in five logical groupings: H Auto Tests H Frequency Checks H Noise Sideband Checks H Amplitude Checks...
Signal generator Frequency range: 1 MHz to 8 GHz; Agilent E8257D with Checking flatness in all bands Output level: - - 30 dBm to +10 dBm; option 1E1 (option UNX Accuracy: <±1 dB recommended) 3- 10 RSA3408B Analyzer Service Manual...
Page 43
4- - 6 DOS-compatible floppy disk correction data for the signal signal generator used) and 4- - 13 generators (Items 1, 2, and 3) 17. Mouse Standard accessary File operation 18. Keyboard Standard accessary File operation 3- 11 RSA3408B Analyzer Service Manual...
Cal > Service... > Password (270833) > PV (see Figure 3- -6). Figure 3- 6: Auto test initial screen The PV menu has the following controls: PV Select. Selects the test items as shown in Table 3- -4. 3- 12 RSA3408B Analyzer Service Manual...
H SG Type. Selects the manufacturer of the generator for the GPIB command system: HP or Anritsu. NOTE. The HP setting also supports Agilent and Rohde & Schwarz products. H GPIB Address. Sets the GPIB address of the generator. 3- 13 RSA3408B Analyzer Service Manual...
Page 46
One SG flatness floppy disk (Item 16) Prerequisites Refer to page 3- - 10 1. Prepare for the test: See the following figure for the hookup. RSA3408B Signal generator GPIB cable Output 10 dB attenuator 50 Ω N-N coaxial cable 3- 14 RSA3408B Analyzer Service Manual...
Page 47
Performance Tests a. Power off the RSA3408B analyzer and the signal generator. b. Connect the analyzer and the generator with a GPIB cable. c. Connect the analyzer Input through a 10 dB attenuator followed by a 50 Ω N-N coaxial cable to the generator output.
One 50 Ω N-N coaxial cable (Item 11) One 10 dB attenuator (Item 14) One GPIB cable (Item 13, two required) One SG flatness floppy disk (Item 16, two required) Prerequisites Refer to page 3- - 10 3- 16 RSA3408B Analyzer Service Manual...
Page 49
H Press the SG Type side key to select the manufacturer of the generator (Item 1) for the GPIB command system: HP or Anritsu. H Press the SG Address side key and set the generator address. H Press the Cancel- -Back (top) side key. 3- 17 RSA3408B Analyzer Service Manual...
Page 50
If the instrument fails any tests, go to When the Auto Tests Fail on page 3- -19. 9. Disconnect the test equipment: a. Disconnect the N-N cable from the analyzer input. b. Turn off the analyzer and generators. c. Disconnect the GPIB cables from all the ports. 3- 18 RSA3408B Analyzer Service Manual...
3. Set the GPIB address of the signal generator: a. Press the SG side key to select the manufacturer of the generator for the GPIB command system: HP or Anritsu. b. Press the GPIB Address side key and set the generator address. 3- 19 RSA3408B Analyzer Service Manual...
Page 52
6. Perform the test: Press the Measurement side key to run the test. Check that the test passes on the right bottom of the screen. 7. Repeat steps 5 and 6 for all the frequencies listed in the log file. 3- 20 RSA3408B Analyzer Service Manual...
Performance Tests Frequency Checks These procedures check the frequency-related characteristics and are listed as checked in Specifications of the RSA3408B Technical Reference. Check Frequency Readout Equipment One signal generator (Item 2 or 3) Required Accuracy One 50 Ω N-N coaxial cable (Item 11) Prerequisites Refer to page 3- - 10.
Page 54
Set the frequency to 2 GHz using the numeric keypad. c. Measure peak frequency: H Press the Peak key to place the marker on the signal peak. H Read the marker readout. Confirm that the value is 2 GHz 1.4 kHz. 3- 22 RSA3408B Analyzer Service Manual...
Page 55
H Press the Peak key to place the marker on the signal peak. H Read the marker readout. Confirm that the value is 7 GHz 2.4 kHz. 6. Disconnect the test equipment: Disconnect the cable at the analyzer input. 3- 23 RSA3408B Analyzer Service Manual...
Page 56
....2. Check the frequency: Check that the frequency counter reads 10 MHz 2.0 Hz. 3. Disconnect the hookup: Disconnect the cable at REF OUT. 3- 24 RSA3408B Analyzer Service Manual...
Performance Tests Noise Sideband Checks These procedures check the noise sideband-related characteristics and are listed as checked in Specifications of the RSA3408B Technical Reference. Check Noise Sideband Equipment One signal generator (Item 4) Required One 50 Ω N-N coaxial cable (Item 11) Prerequisites Refer to page 3- - 10.
Page 58
H Press the Noise Bandwidth side key and set the value to 1 kHz using the numeric keypad. c. Check against limits: Read the phase noise (C/No) at the bottom of the screen. Confirm that the value is 110 dB/Hz or more. 3- 26 RSA3408B Analyzer Service Manual...
Page 59
Check against limits: Read the phase noise (C/No) at the bottom of the screen. Confirm that the value is 132 dB/Hz or more. 5. Disconnect the test equipment: Disconnect the cable at the analyzer input. 3- 27 RSA3408B Analyzer Service Manual...
Performance Tests Amplitude Checks These procedures check the amplitude-related characteristics and are listed as checked in Specifications of the RSA3408B Technical Reference. Check Absolute Equipment One signal generator (Item 1) Required Amplitude Accuracy One power meter (Item 7) One power sensor (Item 8) One 50 Ω...
....25 MHz d. Adjust the output level of the signal generator so that the power meter reads - -10 dBm 0.05 dBm. e. Disconnect the cable from the power sensor input. 3- 29 RSA3408B Analyzer Service Manual...
Page 62
Performance Tests 3. Prepare for the test: a. Hook up the instruments: Connect the signal generator output through a 50 Ω N-N coaxial cable to the RSA3408B Input. See Figure 3- -10. RSA3408B Signal generator Output 50 Ω N-N coaxial cable Figure 3- 10: Hookup for checking the absolute amplitude accuracy b.
Page 63
Hook up the instruments: Connect the signal generator output through a 50 Ω N-N coaxial cable to the RSA3408B Input. See Figure 3- -10 on page 3- -30. b. Modify the analyzer controls: H Press the Frequency/Channel key on the front panel.
Page 64
Press the Peak key on the front panel to place the marker on the signal peak. b. Check that the marker readout is within - -20 dBm 0.5 dB. 9. Disconnect the test equipment: Disconnect the cable at the analyzer input. 3- 32 RSA3408B Analyzer Service Manual...
Input. See the following figure. RSA3408B Signal generator Output 50 Ω N-N coaxial cable c. Initialize the analyzer: H Press the System key on the front panel. H Press the Reset All to Factory Defaults side key. 3- 33 RSA3408B Analyzer Service Manual...
Page 66
Repeat substeps 3a through e for attenuation from 5 to 30 dB in 5 dB steps. 4. Modify the generator controls: Amplitude ....- -5 dBm 3- 34 RSA3408B Analyzer Service Manual...
Page 67
- - P ) is within 0.2 dB. Repeat substeps 7a through e for attenuation from 35 to 55 dB in 5 dB steps. 8. Disconnect the test equipment: Disconnect the cable at the analyzer input. 3- 35 RSA3408B Analyzer Service Manual...
Page 68
....+10 dBm c. Set the power meter control: Frequency ....100 MHz 3- 36 RSA3408B Analyzer Service Manual...
Page 69
... . 10 dB ....Auto 3- 37 RSA3408B Analyzer Service Manual...
Page 70
Signal generator amplitude Reference power - - 40 dBm - - 50 dBm a. Disconnect the cable from the spectrum analyzer. 7. Modify the hookup: a. Connect the signal generator output to the RSA3408B analyzer input. See the following figure. RSA3408B Signal generator Output 50 Ω...
Performance Tests c. Modify the RSA3408B analyzer controls: Center frequency ..100 MHz Span ....
Page 72
Performance Tests Spurious Response Checks These procedures check the spurious-related characteristics and are listed as checked in Specifications of the RSA3408B Technical Reference. Check 3 Order Equipment Two signal generators (Item 2 (or 3) and Item 4) Required Intermodulation Distortion One power combiner (Item 9) Three 50 Ω...
Page 73
H Press the Number Of Averages side key and set the value to 50 using the general purpose knob. H Press the Marker Setup key. H Press the Markers side key to select Delta. 3- 41 RSA3408B Analyzer Service Manual...
Page 74
3. Check against limits: Confirm that the lowest value meets the requirement of - -78 dBc. 4. Disconnect the test equipment: Disconnect the cable at the analyzer input. 3- 42 RSA3408B Analyzer Service Manual...
Page 75
H Press the RBW/FFT key on the front panel. H Press the RBW/FFT side key to select Man. H The RBW menu item is selected by default. Set the value to 10 kHz using the general purpose knob. 3- 43 RSA3408B Analyzer Service Manual...
Page 76
Set the frequency to 3 GHz using the general purpose knob. b. Check against limits: Read the measurement result of Density at the bottom of the screen. Check that the value is - -150 dBm/Hz or less. 3- 44 RSA3408B Analyzer Service Manual...
Page 77
Set the frequency to 7 GHz using the general purpose knob. b. Check against limits: Read the measurement result of Density at the bottom of the screen. Check that the value is - -142 dBm/Hz or less. 3- 45 RSA3408B Analyzer Service Manual...
Page 78
Set the RBW to 100 kHz using the general purpose knob. H Press Trace/Avg > Trace 1 Type... > Average. H Press the Number Of Averages side key and set the value to 50 using the general purpose knob. 3- 46 RSA3408B Analyzer Service Manual...
Page 79
H The Line1 menu item is selected by default. Set the value to - -90 dBm using the numeric keypad. c. Check against limits: Confirm that the residual signal level is - -90 dBm or less. 3- 47 RSA3408B Analyzer Service Manual...
Page 80
Press the Peak key to place the marker at the peak on the noise floor. c. Check against limits: Read the marker readout. Confirm that the residual signal level is - -85 dBm or less within the whole bandwidth. 3- 48 RSA3408B Analyzer Service Manual...
Page 81
....- -5 dBm d. Hook up the signal generator: Connect the generator output through a 50 Ω N-N coaxial cable to the analyzer Input. See the following figure. 3- 49 RSA3408B Analyzer Service Manual...
Page 82
H Press the Peak key to place Marker 2 on the signal peak. H Read the delta marker readout (Δ1- -2) on the screen. Check that the components other than the carrier meet the requirements shown in Table 3- -7 (initially - -73 dBc). 3- 50 RSA3408B Analyzer Service Manual...
- - 73 dBc 2 GHz - - 73 dBc 5 GHz - - 70 dBc 7 GHz - - 70 dBc 3. Disconnect the test equipment: Disconnect the cable at the analyzer input. 3- 51 RSA3408B Analyzer Service Manual...
Page 84
Performance Tests Test Record Photocopy the following test record pages and use them to record the perfor- mance test results for your analyzer. 3- 52 RSA3408B Analyzer Service Manual...
Page 86
- - 50 dB - - 0.2 dB +0.2 dB order intermodulation distortion test Low limit Test result High limit order intermodulation distortion - - 78 dBc +5 dBm reference level, 2 GHz center frequency 3- 54 RSA3408B Analyzer Service Manual...
Page 87
Spurious response test Frequency Low limit Test result High limit 25 MHz - - 73 dBc 2 GHz - - 73 dBc 5 GHz - - 70 dBc 7 GHz - - 70 dBc 3- 55 RSA3408B Analyzer Service Manual...
Page 88
Performance Tests 3- 56 RSA3408B Analyzer Service Manual...
Page 91
Adjustment Procedures This section contains information needed to manually adjust the RSA3408B 8 GHz Real-Time Spectrum Analyzer. The section is divided into three subsections: H General information about adjusting the analyzer H A list of equipment required to perform the adjustments...
Adjustment Procedures Equipment Required Table 4- -1 lists the test equipment required to adjust the RSA3408B analyzer. Table 4- 1: Test equipment Item number and description Minimum requirements Example Purpose Signal generator Frequency range: 10 kHz to 3.5 GHz; Agilent 8648D...
H Store calibration factors in the power meter. H Prepare the flatness correction file for the signal generators. H Prepare the spurious correction file for the signal generators. H Option 03 only. Install the setup files on the AWG520 waveform generator. 4- 3 RSA3408B Analyzer Service Manual...
Page 94
2e9=98.1 4e9=97.1 6e9=96.5 7e9=96.2 8e9=96.0 9e9=95.8 Save the file with the extension of .cal (for example, sensor2.cal). You can load the file to the power meter using a floppy disk or through the network. 4- 4 RSA3408B Analyzer Service Manual...
Press Measure > SG Flatness. Press the Read Cal Factor side key. The file selection dialog box appears. g. Enter the file name and then press the Open button. The calibration factors are stored in the power meter. 4- 5 RSA3408B Analyzer Service Manual...
Page 96
Baseband, RF1, RF2, and RF3. H Procedure 2 on page 4- -10: Uses two signal generators: Item 1, which covers the baseband and RF1 band, and Item 2, which covers the RF2 and RF3 bands. 4- 6 RSA3408B Analyzer Service Manual...
Connect the power sensor output to the power meter input. See Figure 4- -2. RSA3408B Signal generator Power meter GPIB cable GPIB cable Output 10 dB attenuator N adapter N-N coaxial cable Power sensor Figure 4- 2: Initial test hookup 4- 7 RSA3408B Analyzer Service Manual...
Page 98
3. Let the instruments warm up: Allow a 20 minute warm-up period before you go to the next step. 4. Calibrate the power sensor: a. Modify the hookup: Connect the power sensor input to the power reference output of the power meter (see Figure 4- -3). 4- 8 RSA3408B Analyzer Service Manual...
Remove the floppy disk by pressing the eject button. b. Turn off the signal generator, power meter, and analyzer. c. Disconnect the GPIB cables. d. Disconnect the power sensor and the N-N cable. 4- 9 RSA3408B Analyzer Service Manual...
Connect the power sensor output to the power meter input. See Figure 4- -4. RSA3408B Signal generator Power meter GPIB cable GPIB cable Output 10 dB attenuator N adapter Power sensor N-N coaxial cable Figure 4- 4: Initial test hookup 4- 10 RSA3408B Analyzer Service Manual...
Page 101
NOTE. The HP setting also supports Agilent and Rohde & Schwarz products. h. Press the GPIB Address side key and set the address of the generator. 3. Let the instruments warm up: Allow a 20 minute warm-up period before you go to the next step. 4- 11 RSA3408B Analyzer Service Manual...
Remove the floppy disk by pressing the eject button. b. Turn off the signal generator, power meter, and analyzer. c. Disconnect the GPIB cables. d. Disconnect the power sensor and the N-N cable. 4- 12 RSA3408B Analyzer Service Manual...
Page 103
One floppy disk (Item 17) Prerequisites Refer to page 4- - 1 1. Prepare for the calibration: See the following figure for the hookup. RSA3408B Signal generator GPIB cable Output 10 dB attenuator 50 Ω N-N coaxial cable 4- 13 RSA3408B Analyzer Service Manual...
Page 104
Adjustment Procedures a. Power off the RSA3408B analyzer and the signal generator. b. Connect the analyzer and the generator with a GPIB cable. c. Connect the analyzer Input through a 10 dB attenuator followed by a 50 Ω N-N coaxial cable to the generator output.
The auto calibration for the analyzer with Option 03 (IQ input) needs the Waveform Generator AWG520 waveform generator (Item 5). The waveform files must be installed from the AWG520 Setup Disk (floppy disk, Item 18, Tektronix part number (Option 03 Only) 063-3680-00) onto the AWG520 hard disk before executing the calibration.
(see Figure 4- -6). If this happens, run the gain calibration. When UNCAL is displayed, run the gain calibration UNCAL FREE RUN Frequency Span Input Att Figure 4- 6: UNCAL display 4- 16 RSA3408B Analyzer Service Manual...
Cal menu Calibrate all Gain calibration Center offset calibration Cal key DC offset calibration Auto Calibration (Select Yes or No) Figure 4- 7: Calibration menu 4- 17 RSA3408B Analyzer Service Manual...
When you want to perform the gain, center offset, and DC offset calibration together, press the Calibrate All side key. NOTE. Enabling Auto Calibration (see Figure 4- -7) allows the analyzer to run the center offset calibration automatically when analyzer settings change. 4- 18 RSA3408B Analyzer Service Manual...
2. Press the Calibrate DC Offset side key. The calibration runs. It takes several seconds to complete the process. When you want to perform the gain, center offset, and DC offset calibration together, press the Calibrate All side key. 4- 19 RSA3408B Analyzer Service Manual...
Target Files side key Calibration table Shows cal files. Figure 4- 10: Auto calibration initial screen The table on the screen shows cal files for each calibration item. Table 4- -3 summarizes the calibration items. 4- 20 RSA3408B Analyzer Service Manual...
H Other. Selects all the calibration except spurious in the All Band group. All Band. Performs the selected calibration in all bands (Baseband, RF1, RF2, and RF3). Measurement bandwidth: 1 MHz to 8 GHz. 4- 21 RSA3408B Analyzer Service Manual...
Page 112
H SG Address. Sets the GPIB address for the signal generator (Item 1, 2, or 3). H Modulation SG Address. Sets the GPIB address for the modulation signal generator (Item 4). H AWG Address. Sets the GPIB address for the waveform generator (Item 5). 4- 22 RSA3408B Analyzer Service Manual...
Page 113
50 Ω BNC coaxial cable 50 Ω N-N coaxial cable a. Power off the RSA3408B analyzer and the signal generator. b. Connect the analyzer and the generator with GPIB cable. c. Connect the analyzer Input through a 10 dB attenuator followed by a 50 Ω...
Page 114
Press the 10MHz Reference side key to run the calibration. 4. Disconnect the test equipment: a. Disconnect the cable from the analyzer input. b. Turn off the analyzer and generator. c. Disconnect the GPIB cable. 4- 24 RSA3408B Analyzer Service Manual...
Page 115
One SG flatness floppy disk (Item 17) Prerequisites Refer to page 4- - 1 1. Prepare for the calibration: See the following figure for the hookup. RSA3408B Signal generator GPIB cable Output 10 dB attenuator 50 Ω N-N coaxial cable 4- 25 RSA3408B Analyzer Service Manual...
Page 116
Adjustment Procedures a. Power off the RSA3408B analyzer and the signal generator. b. Connect the analyzer and the generator with GPIB cable. c. Connect the analyzer Input through a 10 dB attenuator followed by a 50 Ω N-N coaxial cable to the generator (Item 3) output.
Page 117
10 dB attenuator 50 Ω N-N coaxial cable a. Power off the RSA3408B analyzer and the signal generators. b. Connect the analyzer and the generators with GPIB cables. c. Connect the analyzer Input through a 10 dB attenuator followed by a 50 Ω...
Page 118
Press the RF2+RF3 side key to run the calibration. 6. Disconnect the test equipment: a. Disconnect the N-N cable from the analyzer input. b. Turn off the analyzer and generators. c. Disconnect the GPIB cables from the generators. 4- 28 RSA3408B Analyzer Service Manual...
Page 119
10 dB attenuator 50 Ω N-N coaxial cable a. Power off the RSA3408B analyzer and the signal generator. b. Connect the analyzer and the generator with GPIB cable. c. Connect the analyzer Input through a 10 dB attenuator followed by a 50 Ω...
Page 120
Cal Menu on page 4- -20. 4. Disconnect the test equipment: a. Disconnect the cable from the analyzer input. b. Turn off the analyzer and generator. c. Disconnect the GPIB cable from the generator. 4- 30 RSA3408B Analyzer Service Manual...
Page 121
I- - and Q- - INPUTs coaxial cables a. Power off the RSA3408B analyzer and the waveform generator. b. Connect the analyzer and the generator with a GPIB cable. c. Connect CH1 and CH2 of the generator through 50 Ω BNC coaxial cables to I+ INPUT and Q+ INPUT on the rear panel of the analyzer, respectively.
Page 122
Cal Menu on page 4- -20. 4. Disconnect the test equipment: a. Disconnect the cable from the analyzer input. b. Turn off the analyzer and generators. c. Disconnect the GPIB cables from all the ports. 4- 32 RSA3408B Analyzer Service Manual...
Page 123
....b. Hook up the signal generator: Connect the generator output through a 50 Ω N-N coaxial cable to the analyzer Input. See Figure 4- -11. 4- 33 RSA3408B Analyzer Service Manual...
H With the Span side key, set the span to 3 GHz using the numeric keypad. 3. Check the spectrum: Confirm that a spectrum displays at 1.5 GHz. 4. Disconnect the test equipment: Disconnect the cable at Input. 4- 34 RSA3408B Analyzer Service Manual...
Hook up the signal generator: Connect the generator output through a 50 Ω N-N coaxial cable to the analyzer INPUT. See Figure 4- -12. RSA3408B Signal generator Output 50 Ω N-N coaxial cable Figure 4- 12: Test hookup 4- 35 RSA3408B Analyzer Service Manual...
Page 126
H With the Span side key, set the span to 3 GHz using the numeric keypad. 3. Check the spectrum: Confirm that a spectrum displays at 6.5 GHz. 4. Disconnect the test equipment: Disconnect the cable at Input. 4- 36 RSA3408B Analyzer Service Manual...
Page 129
Maintenance This section contains the information needed to do periodic and corrective maintenance on the RSA3408B 8 GHz Real-Time Spectrum Analyzer. The following subsections are included: H Preparation — Introduction plus general information on preventing damage to internal modules when doing maintenance.
Page 130
5. Handle circuit boards by the edges when possible. 6. Do not slide the modules over any surface. 7. Avoid handling modules in areas that have a floor or work-surface covering capable of generating a static charge. 5- 2 RSA3408B Analyzer Service Manual...
Use only deionized water when cleaning the menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as a cleaner and rinse with deionized water. Before using any other type of cleaner, consult your Tektronix Service Center or representative. 5- 4 RSA3408B Analyzer Service Manual...
CAUTION. To prevent getting moisture inside the analyzer during external cleaning, use only enough liquid to dampen the cloth or applicator. 5- 5 RSA3408B Analyzer Service Manual...
(75% solution) and rinse with warm deionized water. (A cotton- tipped applicator is useful for cleaning in narrow spaces and on circuit boards.) STOP. For further cleaning, contact your local Tektronix service center. 5- 6 RSA3408B Analyzer Service Manual...
Page 135
Cleaning at the beginning of this section.) WARNING. Before doing any procedure in this subsection, disconnect the power cord from the line voltage source. Failure to do so could cause serious injury or death. 5- 7 RSA3408B Analyzer Service Manual...
Page 136
If you are removing a module for service, begin by doing the procedure Access Procedure (page 5- -21). By following the instructions in that procedure, you remove the module to be serviced while removing the minimum number of additional modules. 5- 8 RSA3408B Analyzer Service Manual...
H RF input connector H Rotary encoder H Preamp power H LCD display connector Procedures for H Fan H Board guide bracket 5- - 62 chassis modules H Side panel assembly H Rear panel modules 5- 9 RSA3408B Analyzer Service Manual...
Standard tool Needle-nose pliers Standard tool Retaining ring pliers Standard tool; inch minimum throw Angle-tip tweezers Standard tool Pliers Standard tool Soldering iron Standard tool Solder wick Standard tool Adhesive TRA-CON: Tra-Bond #BA-2114 5- 10 RSA3408B Analyzer Service Manual...
Removal and Installation Procedures Analyzer Orientation Figure 5- -1 shows how the sides of the analyzer are referenced in this manual. DOWN Figure 5- 1: Analyzer orientation 5- 11 RSA3408B Analyzer Service Manual...
Removal and Installation Procedures Rear foot Cabinet Handle Foot Flip stand Line cord Front cushion Foot pad Figure 5- 2: External modules 5- 12 RSA3408B Analyzer Service Manual...
Removal and Installation Procedures Disk bracket Stopper bracket Hard disk drive Shield bracket Shield bracket Floppy disk drive Front connection board CPU board Figure 5- 4: CPU board and disk modules 5- 14 RSA3408B Analyzer Service Manual...
A65 HDD connection board Inside HDD bracket Black sleeving Stopper bracket Cable mount bracket Hard disk drive Shield bracket Back bracket Front bracket Disk bracket Figure 5- 5: Removable hard disk modules (Option 06) 5- 15 RSA3408B Analyzer Service Manual...
Removal and Installation Procedures Power supply 1 Power supply 2 Insulator Shield bracket A71 AC power board Line filter Principal power switch Power supply 3 Figure 5- 7: Bottom modules - Power supply 5- 17 RSA3408B Analyzer Service Manual...
Preamp power A60 Front key board connector Front panel keys Front bezel Rotary encoder Front panel Front bracket Power switch board LCD filter Low front panel General purpose knob Figure 5- 9: Front panel modules 5- 19 RSA3408B Analyzer Service Manual...
- 41 the module A42 DIFP board p.5- - 41 in Figure A70 DC power board p.5- - 42 5- - 6? A50 motherboard p.5- - 44 Figure 5- 11: Guide to removal procedures 5- 21 RSA3408B Analyzer Service Manual...
Page 150
Remove the center two screws securing the cabinet to the rear panel. d. Slide the cabinet backward. As you remove the cabinet, take care not to bind or snag it on the analyzer’s internal cabling. 4. Reinstallation: Reverse step 3 to reinstall the cabinet. 5- 22 RSA3408B Analyzer Service Manual...
Remove the two screws securing the front cushion to the side panel at each side. b. Lifting the front of the analyzer slightly, move the front cushion away. 4. Reinstallation: Reverse step 3 to reinstall the front cushion. 5- 23 RSA3408B Analyzer Service Manual...
Page 152
While holding the handle release, pull the handle out of the slot in the handle cap. Repeat the procedure to remove the handle from the other handle cap. c. Reverse the procedure to reinstall. 5- 24 RSA3408B Analyzer Service Manual...
Page 153
6c; refer to the Cabinet procedure on page 5- -22 to reinstall the cabinet. Reverse steps 3 and 5, reversing each step, to reinstall the flip stand, the handle caps (if removed), and then the handle. 5- 25 RSA3408B Analyzer Service Manual...
Page 154
Removal and Installation Procedures Push down to flex handle and flatten against cabinet then pull it out. Handle cap removal Handle removal Flip stand removal Foot removal Figure 5- 14: Cabinet modules removal 5- 26 RSA3408B Analyzer Service Manual...
Page 155
Locate the RF3 module in the locator diagram RF Modules, Figure 5- -3, page 5- -13. 2. Orient the analyzer: Set the analyzer so the bottom is down on the work surface and its front is facing you. 5- 27 RSA3408B Analyzer Service Manual...
Page 156
Unplug the cable W28 on the RF4 module. c. Remove the two screws securing the module to the main chassis. Lift the module up. 4. Reinstallation: Reverse step 3 to reinstall the RF4 module. 5- 28 RSA3408B Analyzer Service Manual...
Page 157
To REF IN W503 W250 W611 To IF OUT W503 W502 W502 W610 RF5 module RF6 module RF4 module W105 W307 W106 W307 W250 W201 W201 RF2 module RF3 module Figure 5- 15: RF modules removal 5- 29 RSA3408B Analyzer Service Manual...
Page 158
10 MHz REF IN and OUT, W32, W503, W29, W611, W502, and W610. b. Remove the two screws securing the module to the main chassis. Lift the module up. 4. Reinstallation: Reverse step 3 to reinstall the RF5 module. 5- 30 RSA3408B Analyzer Service Manual...
Page 159
CPU board. c. Remove the two screws securing the board to the main chassis. Lift the board up and away from the chassis. 5. Reinstallation: Reverse step 3 and 4 to reinstall the CPU board. 5- 31 RSA3408B Analyzer Service Manual...
Page 160
Removal and Installation Procedures IDE2 FDD1 COM1 CPU board DIMM2 DIMM1 IDE1 COM2 J102 DVO converter CPU board Board stopper bracket Memory support bracket Figure 5- 16: CPU board removal 5- 32 RSA3408B Analyzer Service Manual...
Page 161
Remove the four screws securing the hard disk drive to the disk bracket. b. Lift the hard disk drive up away. 6. Reinstallation: Reverse steps 4 and 5 to reinstall the hard disk drive. 5- 33 RSA3408B Analyzer Service Manual...
Page 162
Stopper bracket Hard disk drive Front connection board Floppy disk drive Shield bracket J122 J120 J112 J114 J110 J900 J106 (FDD) J100 J104 J102 (HDD) A62 Front connection board Figure 5- 17: Disk module removal 5- 34 RSA3408B Analyzer Service Manual...
Page 163
Remove the two screws securing the A62 board to the disk bracket. Lift the board up away. 4. Reinstallation: Reverse steps 2 and 3 to reinstall the A62 board. 5- 35 RSA3408B Analyzer Service Manual...
Page 164
Disconnect the hard disk cable assembly from the hard drive cartridge. Lift the cable assembly up away. 5. Remove the hard drive cartridge: See Figure 5- -18 on page 5- -37. Pull up the hard drive cartridge from the instrument. 5- 36 RSA3408B Analyzer Service Manual...
Page 165
Removal and Installation Procedures Cable mount bracket Black sleeving Hard drive cartridge To IDE1 on the CPU board To J900 on the A62 board Figure 5- 18: Hard drive cartridge removal 5- 37 RSA3408B Analyzer Service Manual...
Page 166
Lift up the bracket assembly. b. Remove the six screws securing the back bracket to the front bracket. c. Remove the two edgings from the top edge of the front bracket. 5- 38 RSA3408B Analyzer Service Manual...
Page 167
10. Remove the A62 Front connection board: See Figure 5- -21. Refer to A62 Front Connection Board on page 5- -35 for the procedure. 11. Reinstallation: Reverse steps 4 through 10 to reinstall the removable hard drive. 5- 39 RSA3408B Analyzer Service Manual...
Page 168
Removal and Installation Procedures Stopper bracket Disk bracket Floppy disk drive Front connection board Shield bracket Figure 5- 21: Disk unit removal 5- 40 RSA3408B Analyzer Service Manual...
Page 169
Locate the A42 DIFP board in locator diagram Motherboard and daughterboards, Figure 5- -6 on page 5- -16. c. Do step 3 in the A10 A/D Board procedure on page 5- -41 to remove the board support bracket. 5- 41 RSA3408B Analyzer Service Manual...
Page 170
Unplug the fan power cable from J300 and J302. b. Unplug the cable W75 (GPIB) from J410. c. Lift up the A70 board. 5. Reinstallation: Reverse step 3 to reinstall the A70 board and interconnect cables. 5- 42 RSA3408B Analyzer Service Manual...
Page 171
Option 03 only W360 Board support bracket W610 W611 Rear support bracket RF6 module A10 A/D board Fan power GPIB J300 J302 J410 A42 DIFP board A70 DC power board Figure 5- 22: Daughter boards removal 5- 43 RSA3408B Analyzer Service Manual...
Page 172
5- -65). c. Reinstall the daughter boards (refer to A10 A/D Board through A70 DC Power Board starting on page 5- -41). d. Reinstall the RF modules (refer to RF Modules on page 5- -27). 5- 44 RSA3408B Analyzer Service Manual...
Page 173
Removal and Installation Procedures Figure 5- 23: Motherboard removal 5- 45 RSA3408B Analyzer Service Manual...
Page 174
Remove the four screws securing the power supply 2 to the bracket. Lift up and remove the power supply 2. 5. Reinstallation: Reverse steps 3 and 4 to reinstall the power supply 1 and 2. 5- 46 RSA3408B Analyzer Service Manual...
Page 175
CN53 W93 to CN1 CN51 CN61 CN81 CN53 CN51 CN61 W84 to CN81 CN61 and CN81 Power supply 1 Power supply 2 Insulator Shield bracket Figure 5- 24: Power supply 1 and 2 removal 5- 47 RSA3408B Analyzer Service Manual...
Page 176
Unplug the cables at CN1 and CN2 on the power supply 3 board. b. Remove the four screws securing the board to the main chassis. Lift up the board away. 4. Reinstallation: Reverse step 3 to reinstall the power supply 3. 5- 48 RSA3408B Analyzer Service Manual...
Page 177
To CN1 on the J110 power supply 3 To J120 on A71 J120 A71 AC power board J130 Line filter Power supply 3 To J100 on A71 Figure 5- 25: Power supply modules removal 5- 49 RSA3408B Analyzer Service Manual...
Page 178
Unplug the cables from the principal power switch. b. Pressing the catches at the top and the bottom of the switch, pull the switch out. 4. Reinstallation: Reverse step 3 to reinstall the principal power switch. 5- 50 RSA3408B Analyzer Service Manual...
Page 179
Locate the step attenuator in the locator diagram Bottom modules - - RF1, Figure 5- -8 on page 5- -18. 2. Orient the analyzer: Set the analyzer so its top is down on the work surface and its left side is facing you. 5- 51 RSA3408B Analyzer Service Manual...
Page 180
J100 (back) J212 J210 J400 J410 J600 J200 To J200 on A100 W122 A100 Cal generator board Step attenuator W120 Attenuator plate To J100 on A100 Figure 5- 26: RF1 modules removal - 1 5- 52 RSA3408B Analyzer Service Manual...
Page 181
Using a #1 Phillips screwdriver, remove the two screws securing the band relay to the bracket. b. Using a inch wrench, disconnect the SMA cables W121, W123, and W124 from the band relay. c. Disconnect the cable W114A from the band relay. 5- 53 RSA3408B Analyzer Service Manual...
Page 182
When removing the A150 YTF interface board, unsolder it at four points. W123 Bandpass filter (YTF) W115 W114 W106 W119 Input relay W120 W101 W121 W114A Band relay W124 W123 Figure 5- 27: RF1 modules removal - 2 5- 54 RSA3408B Analyzer Service Manual...
Page 183
Using an SMA torque driver, disconnect the SMA cables W101, W119, and W120 from the input relay. e. Disconnect the cable W114 from the input relay. 7. Reinstallation: Reverse steps 3 through 6 to reinstall the input relay, band relay, and bandpass filter (YTF). 5- 55 RSA3408B Analyzer Service Manual...
Unplug the cable at J60 on the A50 motherboard. c. Remove the two screws securing the front-panel assembly to the main chassis at each side. Lift the front-panel assembly away. 4. Reinstallation: Reverse step 3 to reinstall the front-panel assembly. 5- 56 RSA3408B Analyzer Service Manual...
Page 185
Using the 12 mm nutdriver, remove the nut securing the rotary encoder to the front bezel. 5. Reinstallation: Reverse steps 3 and 4 to reinstall the rotary encoder and the general purpose knob. 5- 57 RSA3408B Analyzer Service Manual...
Page 186
If servicing the menu buttons, lift them out of the front bezel. 4. Remove the front panel: Unstick the front panel from the front bezel to remove it. 5. Reinstallation: Reverse steps 3 and 4 to reinstall the front panel, menu buttons, and front bezel. 5- 58 RSA3408B Analyzer Service Manual...
Page 187
Remove the one screw securing the A61 board to the front bracket. c. Spreading the catches at left and right sides on the front bracket, lift the A61 board up away. 4. Reinstallation: Reverse step 3 to reinstall the A61 board. 5- 59 RSA3408B Analyzer Service Manual...
Page 188
Locate the LCD display in the locator diagram Front panel modules, Figure 5- -9 on page 5- -19. c. Remove the front-panel assembly (refer to the Front-Panel Assembly procedure on page 5- -56). 5- 60 RSA3408B Analyzer Service Manual...
Page 189
5. Reinstallation: Reverse steps 3 and 4 to reinstall the LCD display and interconnect cables. Front sub-panel Inverter To J710 on the motherboard To J700 on the motherboard RF input connector Figure 5- 30: LCD display removal 5- 61 RSA3408B Analyzer Service Manual...
2. Orient the analyzer: Set the analyzer so its bottom is down on the work surface and its left side is facing you. 3. Perform the CPU Board and Disk Modules procedure on page 5- -31 to remove the CPU board and the disk bracket. 5- 62 RSA3408B Analyzer Service Manual...
Page 191
Pull out the stopper bracket to complete the removal. 5. Reinstallation: Reverse steps 3 and 4 to reinstall the side panel assembly. To J300 on A70 Stopper bracket To J302 on A70 Side panel Figure 5- 31: Fan and side panel removal 5- 63 RSA3408B Analyzer Service Manual...
Page 192
2. Orient the analyzer: Set the analyzer so its bottom is down on the work surface and its left side is facing you. When removing the board guide, press its catches on both sides before pulling it backward. Board guide bracket Figure 5- 32: Board guide bracket removal 5- 64 RSA3408B Analyzer Service Manual...
Page 193
Using a #1 Phillips screwdriver, remove the four screws securing the connector to the rear chassis. Pull the connector away. SMB - - BNC connector Rear panel SMB cable GPIB connector Figure 5- 33: Rear panel removal 5- 65 RSA3408B Analyzer Service Manual...
Page 194
Remove the two screws securing the rear panel to the main chassis. Lift the rear panel up and away from the chassis. 6. Reinstallation: Reverse steps 3 through 5 to reinstall the rear panel. 5- 66 RSA3408B Analyzer Service Manual...
If a fault is detected, first check that the power unit is operating properly according to “Troubleshooting Procedure 1 — Power supply modules”. Then, locate the faulty board according to the troubleshooting tree for each fault or the symptom/cause table. 5- 67 RSA3408B Analyzer Service Manual...
Page 196
(see Table 5- - 7 and Figure 5- - 39). the fan on Replace the fan. the CPU board the voltages working? The power supply is OK. Figure 5- 34: Troubleshooting procedure 1 — Power supply system 5- 68 RSA3408B Analyzer Service Manual...
Page 197
Amplitude: +6.5 dBm ±3 dB Replace RF6 module. Display and CPU system is OK. Replace A42 DIFP board or A50 motherboard Figure 5- 35: Troubleshooting procedure 2 — Display and CPU system 5- 69 RSA3408B Analyzer Service Manual...
Page 198
Amplitude: +6.5 dBm Amplitude: - - 20 dBm A100 Cal generator. ±3 dB ±1 dB Replace Replace RF6 module. Band relay, or A100 Cal generator. Figure 5- 36: Troubleshooting procedure 3 — Signal path 5- 70 RSA3408B Analyzer Service Manual...
Page 199
Replace the A/D board. the spectrum measured as follows? Frequency: 10 MHz ±10 Hz Amplitude: - - 2 dBm ±3 dB Replace RF6 module. Figure 5- 36: Troubleshooting procedure 3 — Signal path (Cont.) 5- 71 RSA3408B Analyzer Service Manual...
Page 200
Replace RF6 module. the spectrum measured as follows? Replace RF6 module. Replace RF4 module. Frequency: 43.125 MHz ±10 Hz Amplitude: - - 2 dBm ±3 dB Figure 5- 36: Troubleshooting procedure 3 — Signal path (Cont.) 5- 72 RSA3408B Analyzer Service Manual...
Page 201
Replace Replace A100 Cal generator board. Step attenuator Input relay or A100 Cal generator. Replace the step attenuator. Figure 5- 36: Troubleshooting procedure 3 — Signal path (Cont.) 5- 73 RSA3408B Analyzer Service Manual...
Page 202
Refer to Reconfiguration of set correctly? RF Modules on page 4- - 33. Replace Restart the analyzer. Bandpass filter or A100 Cal generator. Figure 5- 36: Troubleshooting procedure 3 — Signal path (Cont.) 5- 74 RSA3408B Analyzer Service Manual...
Page 203
W124 W201 W106 To bandpass filter Front connection J700 J810 J710 LCD inverter Power switch J140 J120 J130 Front key J100 Preamplifier power Figure 5- 37: Top view of the instrument for cable interconnection 5- 75 RSA3408B Analyzer Service Manual...
Page 204
J210 W115 J600 J400 J410 J200 To RF3 Bandpass W106 filter (YTF) W114 W119 Step attenuator To RF4 Input relay W120 W101 RF input Figure 5- 38: Bottom view the instrument for cable interconnection 5- 76 RSA3408B Analyzer Service Manual...
CN81 +3.3 V +3.14 V +3.46 V +3.3 V +3.14 V +3.46 V Power supply 3 +5 V +4.75 V +5.25 V +5 V +4.75 V +5.25 V +15 V +14.25 V +15.75 V 5- 77 RSA3408B Analyzer Service Manual...
TP206 +6.6 V +6.27 V +6.93 V Symptoms and Faulty Modules Fault symptoms can help you identify faulty boards or modules. Fault symptoms and suspected boards or modules are listed in Table 5- -8. 5- 78 RSA3408B Analyzer Service Manual...
A42 DIFP board Baseband mode does not work. A10 A/D board RF1 band does not work. RF2 or RF3 module RF2 or 3 band does not work. RF3 module, Bandpass filter, or A100 Cal generator board 5- 79 RSA3408B Analyzer Service Manual...
H You must calibrate IF flatness, RF gain and flatness for RF2 and 3 bands. A100 Cal Generator Board H You must calibrate baseband gain and RF gain for RF1 band. Hard Disk Drive H Full calibration is necessary because the calibration factors are lost. 5- 80 RSA3408B Analyzer Service Manual...
Page 211
Diagrams This section contains the following diagrams: H Block diagram: Shows the modules and functional blocks. H Interconnect diagrams: Show how the modules connect together. 6- 1 RSA3408B Analyzer Service Manual...
Page 212
Diagrams 6- 2 RSA3408B Analyzer Service Manual...
Page 213
GPIB DVO converter CPU board Front connection Front panel key 8.4” LCD Principal power 512 MB memory board board AC power board switch USB X2 ON/STANDBY switch 3.5” FDD Figure 6- 1: Block diagram 6- 3 RSA3408B Analyzer Service Manual...
Page 215
J100 J122 AC power board filter switch J100 J102 J112 DC-DC Voltage Floppy disk Rotary Power switch converter regulator drive encoder board Preamplifier Hard disk Fans power drive GPIB Figure 6- 2: Interconnect diagram 6- 5 RSA3408B Analyzer Service Manual...
Page 219
Replaceable Parts This section contains a list of the replaceable modules for the RSA3408B 8 GHz Real-Time Spectrum Analyzer. Use this list to identify and order replacement parts. Parts Ordering Information Replacement parts are available through your local Tektronix field office or representative.
Page 220
670-7918-03 670-7918-XX When you order parts, Tektronix will provide you with the most current part for your product type, serial number, and modification (if applicable). At the time of your order, Tektronix will determine the part number revision level needed for your product, based on the information you provide.
Items in this section are referenced by figure and index numbers to the exploded view illustrations that follow. Tektronix part number Use this part number when ordering replacement parts from Tektronix. 3 and 4 Serial number Column three indicates the serial number at which the part was first effective. Column four indicates the serial number at which the part was discontinued.
Page 231
Replaceable Parts See Figure 7- - 6 for details. Included in #3. Figure 7- 5: Motherboard and daughterboards 7- 13 RSA3408B Analyzer Service Manual...
Page 232
Name & description number 7- - 6 - - 1 065- - 0776- - XX DISK DRIVE ASSY;PRG 119- - 6832- - XX;RSA3408B W/PACKAGING - - 2 348- - A144- - XX DUMPER - - 3 129- - A593- - XX...
Page 233
Replaceable Parts Figure 7- 6: Disk modules 7- 15 RSA3408B Analyzer Service Manual...
Page 234
GUIDE,CKT BD;NYLON BLACK,19MM L - - 25 407- - 5103- - XX BRACKET,FDD&HDD - - 26 351- - A168- - XX GUIDE,CKT BD;NYLON BLACK,100MM L - - 27 407- - 5102- - XX BRACKET,SHIELD:PC BOARD 7- 16 RSA3408B Analyzer Service Manual...
Page 235
Replaceable Parts See Figure 7- - 8 for details. Black sleeving See Figure 7- - 6- - 1. Figure 7- 7: Disk modules (Option 06) - 1 7- 17 RSA3408B Analyzer Service Manual...
Page 237
Replaceable Parts Figure 7- 8: Disk modules (Option 06) - 2 7- 19 RSA3408B Analyzer Service Manual...
Page 238
CA ASSY,SP,ELEC:2,18AWG,13CM L,W/FASTON TO HLDR 3P,W/TUBE - - 24 174- - C005- - XX CA ASSY,SP,ELEC:2,18AWG,15CM L,W/HOLDER,HLDR 5P TO 3P - - 25 174- - C006- - XX CA ASSY,SP,ELEC:2,18AWG,9CM L,W/HOLDER,HLDR 5P TO 3P 7- 20 RSA3408B Analyzer Service Manual...
Page 239
Replaceable Parts Figure 7- 9: Bottom modules - Power supply 7- 21 RSA3408B Analyzer Service Manual...
Page 240
348- - 0949- - XX GROMMET,PLASTIC:NYLON,BLACK,RING,15MM ID - - 25 174- - B949- - XX CA ASSY,RF:50OHM,SEMI- - RIGID,W106,FINE ATT TO SW - - 26 174- - C041- - XX CABLE ASSY,RF:SEMI- - RIGID,W124,SW TO A310 7- 22 RSA3408B Analyzer Service Manual...
Page 241
Replaceable Parts A150 A100 Figure 7- 10: Bottom modules - RF1 7- 23 RSA3408B Analyzer Service Manual...
Page 242
211- - 1028- - XX SCREW,MACHINE:M4X8MM L,PNH,STL,ZN- - C PL,CROSS REC W/FLAT&LOCK WASHER - - 18 441- - A293- - XX CHASSIS,MAIN - - 19 211- - 1028- - XX SCREW,MACHINE:M4X8MM L,PNH,STL,ZN- - C PL,CROSS REC W/FLAT&LOCK WASHER 7- 24 RSA3408B Analyzer Service Manual...
Page 244
SHLD GSKT ELEC:CONDUCTIVE URETHANE FOAM,1X4 W/ADHESIVE,10.5CM - - 11 348- - A158- - XX SHLD GSKT ELEC:CONDUCTIVE URETHANE FOAM,1X4 W/ADHESIVE,10CM - - 12 348- - A143- - XX GASKET,SHIELD:FINGER TYPE,15.2MM W X 5.6MM H X 7.2MM L 7- 26 RSA3408B Analyzer Service Manual...
Page 245
Replaceable Parts Figure 7- 12: Chassis modules - 2 7- 27 RSA3408B Analyzer Service Manual...
Page 246
- - - - - - - - - - - - - - - - SCREW 10- - 32X0.354INCH,TRUSS (Included in the slide assembly (Figure index 10- - 13- - 7)) - - 13 335- - 0391- - XX MARKER,IDENT;SAFETY LABEL 7- 28 RSA3408B Analyzer Service Manual...
Page 247
Replaceable Parts Figure 7- 13: Rackmount 7- 29 RSA3408B Analyzer Service Manual...
Page 248
MANUAL,TECH:PROGRAMMER,ENGLISH,PDF ONLY, INCLUDED ON 063- - 4089- - XX SOFTWARE CD OPTIONAL ACCESSORIES 071- - 2366- - XX MANUAL,TECH:SERVICE,ENGLISH 071- - 2480- - XX MANUAL,TECH:TECHNICAL REFERENCE;PDF,ENGLISH 063- - 3680- - XX SOFTWARE PKG:3.5FD,AWG520 SETUP DISK 7- 30 RSA3408B Analyzer Service Manual...