Summary of Contents for Tektronix ECO422D
Page 1
Instruction Manual ECO422D SD/HD Changeover Unit 071-0859-01 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Page 3
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation. Symbols and Terms Terms in this Manual. These terms may appear in this manual: ECO422D SD/HD Changeover Unit Instruction Manual...
Page 12
CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product: Double Protective Ground CAUTION WARNING Refer to Manual High Voltage Insulated (Earth) Terminal viii ECO422D SD/HD Changeover Unit Instruction Manual...
Page 13
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components. To avoid electric shock, do not touch exposed connections. ECO422D SD/HD Changeover Unit Instruction Manual...
Page 14
Service Safety Summary ECO422D SD/HD Changeover Unit Instruction Manual...
Page 15
H Specifications describes functional characteristics and performance require- ments for the ECO422D. H Theory of Operation explains how the ECO422D operates to a level sufficient to perform component level service. H Performance Verification describes how to verify the functional performance of the ECO422D.
Page 16
This phone number is toll free in North America. After office hours, please leave a voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices. ECO422D SD/HD Changeover Unit Instruction Manual...
Getting Started Product Description The ECO422D SD/HD Changeover Unit (Figure 1- -1 and Figure 1- -2) provides automatic selection of reference sources. Automatic changeover may occur upon fault detection in any active source. Automatic transfer ensures uninterrupted signals for critical applications. The ECO422D is an in-line device without internal buffers.
17-¾ inches to allow clearance for the slide-out tracks. Rack slides conveniently mount in any rack that has a front-to-rear rail spacing between 15-½ and 28 inches. The ECO422D requires six inches of clearance between the instrument rear panel and any rear cabinet panel for connector space and to provide adequate air circulation.
Getting Started Bar nut (Use if the front rail is not tapped) Figure 1- 3: Front rail mount Bar nut screws Rear rack rail Figure 1- 4: Deep rackmount 1- 3 ECO422D SD/HD Changeover Unit Instruction Manual...
(The right hand rail is Flat nut bar shown above.) 10-32 PHS screw Stop latch hole Intermediate section Figure 1- 6: Assembly of rackmounting hardware 1- 4 ECO422D SD/HD Changeover Unit Instruction Manual...
1. Loosen retaining screw and pull instrument outward until the stop latches snap into the holes. 2. Press stop latches and remove instrument. Stop latch Figure 1- 7: Installing or removing the rack slides 1- 5 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 24
Electrical Installation You can order any of the following power cord options for the ECO422D. If no power cord option is ordered, the instrument is shipped with a North American 125 V power cord and one replacement fuse.
Figure 1- -2 shows the instrument rear panel. Connecting Power To supply power to the ECO422D, connect the power cord to the power-input connector on the rear-panel of the instrument, and then connect the power cord to the local mains supply.
(example: Trompeter RT- -1L). There are many different ways to configure the ECO422D. The illustration below is only one example. One thing to keep in mind, the ECO422D will automatically change to the Primary signals whenever it looses power. Therefore, always power the Backup source and the ECO422D by the same source and the Primary source separately.
Page 27
ECO422D. This section gives one example, analog active video. Active Analog Video. The ECO422D will pass active analog video by using the no checking mode, but there will be times when you need to check for the presence of an active video signal.
All of these switches are available through a small access panel on top of the ECO422D, eliminating any need to remove the top cover just to change the function of the outputs.
Getting Started Functional Check and First Time Operation This section steps through how the ECO422D will react under its most common operating circumstances: two good signals, loss and then return of the Primary signal, loss and return of the Backup signal, and loss of power. Only two input signals are used here to avoid confusion.
Page 31
Power up all instruments. Enable the front panel, using the Front Panel button. Set the ECO422D to “Switch on Fault” using front-panel buttons. (This is the default at power on.) If Primary is not already the source, press the Sync Source button to make Primary the source.
Page 32
Getting Started Reconnect the Primary input source. (Replace the cable between the SPG422 and the Primary 1 input of the ECO422D.) Press the Reset button to clear the fault indicator LED. Return the auto switching to “Switch on Fault.” Note that the sync source is still Backup.
Primary or Backup input). It can be manually changed using the Sync Source button. (If the ECO422D is in Normal mode, this is only true when the other input is good or both inputs are bad. Refer to User Configuration Switch (S14- -8) on page 2- -3.) Only one of these LEDs can be on.
Serial digital video (270 Mb/s) Analog tri- - level sync NTSC serial digital audio User defined User defined Serial digital video (1.485 Gb/s) (Channels 1- - 6) Attenuation (Channels 7- - 11) 2- 2 ECO422D SD/HD Changeover Unit Instruction Manual...
For Normal operation, the user cannot switch to a bad signal whether the instrument is in manual or auto switch mode. For example, the ECO422D is in manual mode and the Primary signal is bad, while the Backup signal is good. If the user presses the Sync Source button, the output will be the Backup signal.
Sync signal. The procedure used to set these levels is on page 2- -8. Connectors Figure 2- -3 shows the layout of the rear panel of the ECO422D. Figure 2- 3: Rear panel of the ECO422D Power This instrument is designed to operate from a single-phase power source with one current-carrying conductor at or near earth-ground (the neutral conductor).
Operating Basics Output. Signal output. It is either from the Primary or the Backup source. How the ECO422D is configured determines under what conditions the source changes. Backup. Input from the backup sync generator. This should be the same signal type as its Primary signal pair.
Fault Reporting - - and + It is normally open, indicating that everything is good. It will close in the cases of: loss of power to the ECO422D or one (or more) input signals are bad. The fault is latched and will remain until the fault is cleared.
State Machine Description Table 2- -3 is a truth table that gives the various states the ECO422D outputs based on the state of the inputs. Note that Primary is set to check a given level.
Operating Basics Table 2- 3: Truth table for ECO422D switching Power User config Switch O auto O manual " " " " " " " " " " " " " " " " " " How to Adjust the...
3. Set all S14 switches “closed.” 4. Set the ECO422D to Primary Sync Source. 5. Set the ECO422D Auto Switching to “Switch on Fault.” 6. Set the step attenuator to 0 dB of attenuation. 7. Turn the potentiometer R265 fully. (You are attempting to allow the minimum signal level to pass through before the threshold is reached.)
Operating Basics 12. Slowly adjust R265 until the ECO422D switches to the Backup sync source and a fault occurs on the Primary channel. Press Reset to verify that the fault is still preset. If the Primary fault indicator LED goes out, continue adjusting the potentiometer.
Specifications Introduction The items listed in the following tables describe the performance of the ECO422D Changeover Unit. Performance Requirements are generally quantita- tive and can be tested by a Performance Verification procedure contained in the service part of this manual.
Time that it takes for the relays to switch and settle. Approximately 10 msec. Power-up delay > 4 minutes, to allow source generator signals to stabilize Approximates less than 20 meters of Belden 1694A cable. 3- 2 ECO422D SD/HD Changeover Unit Instruction Manual...
Attenuation approximately x5 Analog tri-level sync 400 to 565 mV 475 mV (between 3 and 6 dB down) Performance is verified indirectly by completing Performance Verification steps 2- 5 & 10- 11. 3- 4 ECO422D SD/HD Changeover Unit Instruction Manual...
Class I (grounded product), as defined in IEC 1010-1, Annex H. Installation Category Installation Category II, as defined in IEC 1010-1, Annex J. Rated for indoor use only. Pollution Degree Pollution Degree 2, as defined in IEC 1010-1. 3- 6 ECO422D SD/HD Changeover Unit Instruction Manual...
FCC Compliance Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits. High quality shielded cables must be used to ensure compliance to the above listed standards. 3- 7 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 54
CAT II Local-level mains (wall sockets). Equipment at this level includes appliances, portable tools, and similar products. Equipment is usually cord-connected. CAT I Secondary (signal level) or battery operated circuits of electronic equipment. 3- 8 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 55
Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind. 3- 9 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 57
WARNING The following servicing instructions are for use only by qualified personnel. To avoid injury, do not perform any servicing other than that stated in the operating instructions unless you are qualified to do so. Refer to all safety summaries before performing any service.
Theory of Operation This section provides information on how the ECO422D circuitry works, in order to troubleshoot the instrument. Logic Conventions Signal names are all capital letters. For example, SIGNAL. An active low signal (normally denoted by an overscore) is a signal name enclosed in parentheses.
Latch Flag other channels Figure 4- 1: Block diagram of the ECO422D The Primary and Backup signals enter through the BNCs. The signal goes directly to the Relay where one switches through to the rear-panel Output and the other terminates in 75 Ω. The rest of the circuitry determines which signal goes to the Output.
This allows checking a broader range of signals without the Negative Peak Detectors being overdriven or saturated. The output of the Input Buffers goes to the Negative Peak Detectors. 4- 3 ECO422D SD/HD Changeover Unit Instruction Manual...
Controller PLD that then generates the control signals for the switching relays. (P_AND) (P_AND_C) (B_AND) (B_AND_C) Keyboard Remote encoder Controller Timer/ flasher Front panel led driver Watchdog timer Figure 4- 3: Block diagram for the Control and Remote circuit 4- 4 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 65
Q33 to saturate to less than 1.5 V, signaling the fault. When the LED is on, Q33 is cut off leaving less than 100 nA of leakage. The diodes, CR48 and VR1, are provided for reverse voltage and ESD protec- tion. 4- 5 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 66
(P_SELECT) and (B_SELECT) signals. (AUTO_ON) and (AUTO_OFF) are sent to drive the front-panel LED circuit. After the front panel is enabled, the ECO422D will only allow the front panel approximately 60 seconds of inactivity before it will automatically disable the front panel.
ATTEN, is inserted. Opening the seven reference amplitude switches allows VREF to go to a negative voltage. Since the peak detector will never produce a negative voltage, the comparators will be disabled. 4- 7 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 68
The replacement fuse on the power supply should be a 3.15 A, 250 V fuse. The rear panel fuse requirements are listed in Replaceable Electrical Parts. 4- 8 ECO422D SD/HD Changeover Unit Instruction Manual...
In addition, it can be used as a short-form Performance Check for those familiar with the Performance Verification Procedure. Only steps that have numeric Performance Requirements are included in this report form. (Some steps have been omitted.) 5- 1 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 72
5 dB down Backup NTSC Serial Digital Composite switch at 2 dB down switch at 5 dB down Switch - - Primary Certificate number is not provided, unless “Certificate of Traceability” is issued. 5- 2 ECO422D SD/HD Changeover Unit Instruction Manual...
Serial Digital 270 Mb/s Switch - - switch at 2 dB down switch at 5 dB down Backup NTSC Serial Digital Composite switch at 2 dB down switch at 5 dB down Switch - - Primary 5- 3 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 74
Serial Digital 270 Mb/s Switch - - switch at 2 dB down switch at 5 dB down Backup NTSC Serial Digital Composite switch at 2 dB down switch at 5 dB down Switch - - Primary 5- 4 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 75
Serial Digital Audio Switch - - Primary switch at 2 dB down switch at 4 dB down Serial Digital Audio Switch - - Backup switch at 2 dB down switch at 4 dB down 5- 5 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 76
PAL Black Burst Switch - - Backup switch at 2 dB down switch at 4 dB down Serial Digital 270 Mb/s Switch - - switch at 2 dB down switch at 5 dB down Primary 5- 6 ECO422D SD/HD Changeover Unit Instruction Manual...
1 GHz bandwidth, for example, a TDS7000 Series oscilloscope; Example waveform monitor: 1745A Waveform Monitor.) 75 Ω terminators. (Example: Tektronix part number 011-0163-00 for serial digital video or 011-0102-01 for analog and audio) 75 Ω Cables. (Example: Tektronix part number 012-0074-00)
Set all switches on S1 through S11 to “open” to disable switches on all channels. Power the ECO422D up. (No inputs or outputs are required at this point.) Wait for the power-up delay to complete (over 4 minutes). CHECK - - that there are no error shown on the front panel.
Page 79
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D to “Primary” Sync Source and Auto Switch “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Primary input signal until the signal source switches to the Backup.
Page 80
Primary. CHECK - - that the Primary signal still shows an error and automatically switches back to Backup. If the ECO422D does not continue to show the Primary signal in error after the reset, continue adding attenuation until the Primary signal switches to Backup again.
Page 81
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D to “Backup” Sync Source and Auto Switch to “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Backup input signal until the signal source switches to the Primary.
6. Check the NTSC & PAL Black Burst for Channels 2 - - 11 Switch between 2 and 4 dB down from nominal Repeat the four previous procedures for Channels 2 - - 11 of the ECO422D. Move the cables from Channel 1 to the channel under test. Replace the switch setting for S1 as given in the following Table.
Page 83
Primary. CHECK - - that the Primary signal still shows an error and automatically switches back to Backup. If the ECO422D does not continue to show the Primary signal in error after the reset, continue adding attenuation until the Primary signal switches to Backup again.
Page 84
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D (using the front panel) to “Backup” Sync Source and Auto Switch “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Backup input signal until the signal source switches to Primary.
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D to “Primary” Sync Source and Auto Switch to “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Primary input signal until the signal source switches to Backup.
Page 86
Performance Verification CHECK - - that the Primary signal still shows an error and automatically switches back to Backup. If the ECO422D does not continue to show the Primary signal in error after the reset, continue adding attenuation until the Primary signal switches to Backup again.
Switch between 2 and 5 dB down from nominal Repeat the previous two procedures for Channels 2 through 11 of the ECO422D. Move the cables from Channel 1 to the channel under test. Replace the switch setting for S1 as given in the following table.
Page 88
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D to “Primary” Sync Source and Auto Switch to “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Primary input signal until the signal source switches to Backup.
Page 89
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D to “Backup” Sync Source and Auto Switch to “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Backup input signal until the signal source switches to Primary.
Page 90
CHECK - - that there are no errors shown for either the Primary or Backup ECO422D sync sources. Set the ECO422D (using the front panel) to “Primary” Sync Source and Auto Switch “Switch on Fault.” Add attenuation, 0.5 dB at a time, to the Primary input signal until the signal source switches to Backup.
Page 91
Performance Verification CHECK - - that the Primary signal still shows an error and automatically switches back to Backup. If the ECO422D does not continue to show the Primary signal in error after the reset, continue adding attenuation until the Primary signal switches to Backup again.
Page 95
Adjustment Procedure The only adjustment required for the ECO422D is the user-defined threshold levels. Adjustment Procedure How to Adjust the The user-defined threshold levels are the two levels available from User User-Defined Threshold Configuration switches SX- -6 (R266) and SX- -7 (R265). They are available so that the user can check the levels of signals other than the five predefined levels.
Inspection and Cleaning Preventive maintenance consists of cleaning, lubricating, visual inspection, checking performance, and readjusting the ECO422D on a regular schedule. The Inspection and cleaning schedule should be established based on the amount of use and the surrounding environment of the ECO422D. Under average condi- tions, a preventive maintenance check should be performed on a one year interval (severe environmental conditions may dictate a shorter time interval).
Page 100
Periodic checks of the transistors and integrated circuits are not recommended. The best measure of performance is the actual operation of the component in the circuit. 7- 2 ECO422D SD/HD Changeover Unit Instruction Manual...
Service static-sensitive components or assemblies only at a static free workstation by qualified personnel. If soldering is involved, use a soldering iron connected to earth ground and special antistatic desoldering tools. 7- 3 ECO422D SD/HD Changeover Unit Instruction Manual...
Ensure that the malfunction exists in the instrument. This is done by making sure Procedures that the instrument is operating as intended by Tektronix (see Operating Basics), and by checking that a malfunction has not occurred upstream from the ECO422D.
Page 103
Determine the extent of the repair. If the necessary repair is complex, it may be advisable to contact your local Tektronix field office or representative before continuing. If the repair is minor, such as replacing a component, see the parts list for replacement information.
In addition, there are training programs that are available for service technicians. NOTE. When considering which service offerings best suit the current need, remember that Tektronix provides a limited parts and service warranty for all its products. No customer repairs should be attempted during the warranty period. Service Training Tektronix provides service training in a number of programs.
Note that this information is important whether you are ordering a circuit board from the Module Exchange Center or as a new replacement part from Tektronix. Table 7- 2: Assemblies in the ECO422D Circuit board First seven digits Assembly name assembly No.
Page 106
Maintenance Factory Replacement Replacement parts are available through the local Tektronix field office or Parts representative. However, many common electric parts are available through local sources. Using a local source, where possible, will eliminate shipping delays. Changes to Tektronix instruments are sometimes made to accommodate improved components, as they become available, and to improve circuit performance.
NOTE. During installation, remember to include the spacer strip on the BNC connectors. Remove these screws Remove these posts Figure 7- 1: Remove these screws and posts before attempting to remove the Connector board 7- 9 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 108
Figure 7- 2: Slide the Connector board out this way CAUTION. Use extreme care when sliding any of the boards to clear the rear panel because there are components on both sides of the boards. 7- 10 ECO422D SD/HD Changeover Unit Instruction Manual...
Repackaging Instructions Identification Tag If the instrument is to be shipped to Tektronix Service Center for service or repair, attach a tag to the instrument showing: 1. Owner (with complete address) and the name of the person at your firm that can be contacted.
Replaceable Electrical Parts This section contains a list of the components that are replaceable for the ECO422D. Use this list to identify and order replacement parts. There is a separate Replaceable Electrical Parts list for each instrument. Parts Ordering Information Replacement parts are available from or through your local Tektronix, Inc., Field...
These mechanical subparts are listed with their associated electrical part (for example, fuse holder follows fuse). Chassis-mounted parts and cable assemblies have no assembly number prefix and are located at the end of the electrical parts list. 8- 2 ECO422D SD/HD Changeover Unit Instruction Manual...
Indicates the code number of the actual manufacturer of the part. (Code to name and address cross reference can be found immediately after this page.) (Column 6) Mfr. Part No. (Column 7) Indicates actual manufacturer’s part number. 8- 3 ECO422D SD/HD Changeover Unit Instruction Manual...
SAN JOSE CA 95131- - 1008 OPTOELECTRONICS DIV 53387 3M COMPANY 3M AUSTIN CENTER AUSTIN TX 78769- - 2963 ELECTRONIC PRODUCTS DIV 55680 NICHICON /AMERICA/ CORP 927 E STATE PKY SCHAUMBURG IL 60195- - 4526 ECO422D SD/HD Changeover Unit Instruction Manual 8- 4...
Page 117
PO BOX 500 81073 GRAYHILL INC 561 HILLGROVE AVE LA GRANGE IL 60525- - 5914 PO BOX 10373 91637 DALE ELECTRONICS INC 2064 12TH AVE COLUMBUS NE 68601- - 3632 PO BOX 609 ECO422D SD/HD Changeover Unit Instruction Manual 8- 5...
Page 139
Example: ID CONTROL, (ID CONTROL), or /ID CONTROL. Abbreviations are based on ANSI Y1.1- -1972. Other ANSI standards that are used in the preparation of diagrams by Tektronix, Inc. are: H Y14.15, 1966 — Drafting Practices. H Y14.2, 1973 — Line Conventions and Lettering.
Page 140
---15V SEE PARTS LIST FOR CIRCUIT BOARD EARLIER VALUES AND OUTLINE SERIAL NUMBER RANGES. PART OF A1 MAIN BOARD INSTRUMENT NAME VERTICAL INPUT SCHEMATIC NAME & NUMBER ASSEMBLY NUMBER COMPONENT LOCATOR GRID 9- 2 ECO422D SD/HD Changeover Unit Instruction Manual...
Page 141
A1 Front Panel Board Static Sensitive Devices See Maintenance Section Front Panel Board and Schematic Diagram <1> Component Locator Chart A1 Front Panel Board. Comp Diag Comp Diag ECO422D...
Page 142
SYNC SOURCE AUTO SWITCHING FAULT INDICATOR FRONT PANEL PRIMARY SWITCH ON FAULT PRIMARY ENABLED ENABLE P_FAULT AUTO PRIMARY ENABLE DISABLE P_FAULT B_FAULT BACKUP DISABLED BACKUP LOCKED OUT AUTO CONNECTS PRIMARY TO J16 DISABLE BACKUP B_FAULT BACKUP MAIN RESET BOARD UNUSED PUSHBUTTON LED’S MOUNTING HOLES A1 FRONT PANEL BOARD ECO 422 Changeover Unit...
Page 157
Replaceable Mechanical Parts This section contains a list of the components that are replaceable for the ECO422D. Use this list to identify and order replacement parts. There is a separate Replaceable Mechanical Parts list for each instrument. Parts Ordering Information Replacement parts are available from or through your local Tektronix, Inc., Field...
Page 158
Indicates the code number of the actual manufacturer of the part. (Code to name (Column 7) and address cross reference can be found immediately after this page.) Mfr. Part Number Indicates actual manufacturer’s part number. (Column 8) ECO422D SD/HD Changeover Unit Instruction Manual 10- 2...
Page 159
BEAVERTON OR 97077- - 0001 PO BOX 500 80126 PACIFIC ELECTRICORD CO 747 W REDONDO BEACH GARDENA CA 90247- - 4203 PO BOX 10 85471 BOYD CORP 13885 RAMOMA AVE CHINO CA 91710 ECO422D SD/HD Changeover Unit Instruction Manual 10- 3...
This instrument is orderable with an “A” option to accommodate the various national electrical power connections. Power Cord Options Any of the following power cord options can be ordered for the ECO422D. If no power cord option is ordered, instruments are shipped with a North American 125 V power cord.
Table A- 1: Power cord identification Plug configuration Normal usage Option number North America Standard 125 V/15 A Plug NEMA 5-15P Europe 230 V United Kingdom 230 V Australia 230 V Switzerland 230 V A- 2 ECO422D SD/HD Changeover Unit Instruction Manual...