Running a Control Solution Test with EasyMax 15 Control Solution
You need the meter, a test strip, and control solution
1. Put a test strip into the meter
in the direction of the arrow.
Note: If the Patient Medical Record is on, it does not matter which record number you select if you
flag the result as a control solution test. The result will be displayed in the meter's memory
as a control solution test.
6. Touch the drop to the blood
collection area at the tip of
the test strip. Do not put
control solution on the top
of test strip. The meter
starts to count down from
5 seconds and will show the
results.
2. Press
(Left Button) to
select the mode of
Control Solution, and the
icon of flashes.
7. Do not remove the test
strip until you confirm that
the reading falls within the
range printed on the test
strip vial. Be sure to match
the correct control solution
range to your test result.
Page 7
3. Place the meter on a flat
surface, like a table.
4. Remove the control solution
bottle cap and wipe the tip
of the bottle with a tissue.
5. Squeeze the bottle until a
tiny drop forms at the tip of
the bottle and place
solution on top of the bottle
cap.
8. Push the Strip Ejector to
eject the test strip into a
proper trash receptacle.