Section 2
Operation
2-1.
INTRODUCTION
2-2,
This
section describes
how
to
set
up and
make
measurements
with
your
Multimeter.
Even
though
you
may
have used a multimeter
before,
we recommend
that
you
read
the
entire
section carefully
so
that
you
can
use
all
of the
features
of
your
Multimeter.
2-3.
SETTING
UP
YOUR
INSTRUMENT
2-4,
Unpacking
2-5.
Your
Multimeter
is
shipped
in
a
special protective
container that
should
prevent
damage
to
the
instrument
during
shipping.
Check
the
shipping order
against the
contents of the container
and
report
any
damage
or short
shipment
to
the place of
purchase
or the nearest
Fluke
Technical Service Center.
A
list
of these service centers
is
located
in
Section
5.
The
container should include
the
following:
The 8010A
or
8012A
Multimeter
Two
test
leads
(one
red
and one
black)
Line
power
cord
The 8010A/8012A
Instruction
Manual
2-6.
If
reshipment of
the
instrument
is
necessary, please
use the
original
shipping
container.
If
the original
container
is
not
available,
be
sure that
adequate
protection
is
provided
to
prevent
damage
during
shipment.
We
recommend
that the
instrument be
surrounded by
at
least
three inches of
shock-absorbing
material
on
all
sides
of the
container.
2-7.
Remove
the
Multimeter
from
its
container
and
place
it
in
a convenient
location.
The
carrying
handle
on
the
meter can be used
as
a prop-stand
or
positioned out of
the
way
(behind
the
Multimeter).
To
position the handle,
pull
outward
on
the
hubs
of the
handle
and
rotate the
handle
into position.
2-8,
AC
Line Voltage
Requirements
2-9.
AC
line
voltage
requirements
for
your Multimeter
are
listed
on
a
decal attached to the
bottom
of the
instrument.
Refer
to
Section 4 for the procedure
to
change
the ac
line
voltage
setting.
If
your Multimeter has
the
-
01 Battery
Option,
refer to
Section 6
for
information
on
changing
ac
line
voltages.
CAUTION
Do
not
connect
the
power
cable
to
the
instrument before
verifying that
the intended
source
matches
the
ac
line
configuration
of
the instrument.
2-10.
Fuse Replacement
2-11.
There
is
one, user-replaceable fuse (FI)
in
your
Multimeter.
The
fuse (FI)
and
the fuse
holder
form an
integral
part
of
the
mA
input
connector
and can be
removed
with ordinary
tools.
The
fuse rating
is:
2
A,
normal blow (recommended
part
AGX2).
2-12.
Use
the
following
procedure
to replace the
fuse,
FI:
1.
Set the
POWER
switch to
OFF.
2.
Remove
the
input
power
cord
from
the
Multimeter.
WARNING
DO NOT
REPLACE THE FUSE WITH THE
INSTRUMENT
TURNED ON OR CONNECT-
ED
TO
LINE
POWER.
3.
Refer
to
Figure
2-1,
item
5
for the location of
the fuse holder.
2-1