Fluke 8010A Instruction Manual page 6

Hide thumbs Also See for 8010A:
Table of Contents

Advertisement

INTRODUCTION
AND
SPECIFICATIONS
SPECIFICATIONS
Table
1-2.
8010A/8012A
Specifications (cont)
Resistance
RANGE
RESOLUTION
ACCURACY:
for
1-Year
FULL-SCALE
VOLTAGE
MAXIMUM
TEST
CURRENT
200
n
o.m
±(0.2%
of reading
+1
digit)
<
0.25V
1
.3
mA
>
1.0V
1
.3
mA
20
kD
10D
<
0.25V
IOmA
200
kD
^
100D
>
uov
35
mA
2000
kn
1
kn
±(0.5%
of reading
+^
digit)
<
0.25V
0.10
juA
20
MD
-M-
10
kD
>
1.5V
0.35
idA
OVERLOAD PROTECTION
300V
dc/ac
rms on
all
ranges.
O'PEN
CIRCUIT
VOLTAGE
Less than 3.5V
on
all
ranges,
RESPONSE
TIME
1
second
all
ranges except 4
seconds on
the
2000
kO
and 20
MO
ranges,
DIODE TEST
These
three
ranges have
enough
voltage
to
turn
on
silicon
-W-
junctions
to
check
for
proper forward-to-back
resistance.
The
2
kO
range
is
preferred
and
is
marked
with a
larger
diode
symbol on
the
front
panel
of
the instrument.
The
three
non-diode
test
ranges
will
not turn
on
silicon
junctions so
in-circuit
resistance
measurements can
be
made
with
these three ranges.
Low
Resistance 801
2A
Only
RANGE
RESOLUTION
ACCURACY:
for
1-Year
FULL-SCALE
VOLTAGE
MAXIMUM
TEST
CURRENT
2D
(LO Ohms)
1
mD
±(1%
of reading
+2
digits)
0.02V
1
0.5
mA
20D
(LO Ohms)
10
mD
±(0.5%
of
reading
+2
digits)
0.2V
10.5
mA
OVERLOAD
PROTECTION
300V
dc/ac
rms on
ail
ranges.
RESPONSE
TIME
1
second
maximum.
OPEN
CIRCUIT
VOLTAGE
16V
maximum
on
both
ranges.
Conductance
RANGE
RESOLUTION
ACCURACY;
for
1-Year
OPEN
CIRCUIT
VOLTAGE
MAXIMUM
TEST
CURRENT
2
mS
1
mS
±(0,2%
of
reading
+
1
digit)
<3.5V
1.3
mA
20
mS
10
nS
<1V
IOmA
200 nS
.1
nS
±(1%
of reading
-t-10
digits)
<1V
0.10
mA
OVERLOAD
PROTECTION
300V
dc/ac
rms on
ail
ranges,
CONDUCTANCE
UNITS
We
use
the
international unit of
conductance,
the
Siemen
=
S
=
1/D.
Another
unit of
conductance
is
the
mho.