MAINTENANCE
PERFORMANCE
TEST
Figure
4-3.
Calibration
Adjustment
Locations
4-44.
AC
Current Test
445.
Use
the following
procedure
to verify
the
proper
operation
and
accuracy of
the
DC
Current
function:
1
Connect
the
HI
output of
the
DMM
Calibrator
to
the
mA
input terminal
of the
UUT.
Connect
the
LO
output of
the
DMM
Calibrator to the
COMMON
input terminal of the
UUT.
Table
4-5.
AC
Voltage Test
STEP
RANGE
UUT
INPUT
LEVEL
FREQ
190.0
mV
100
Hz
10.0
mV
100 Hz
1
.900V
100
Hz
19.00V
100
Hz
190.0V
100
Hz
750V
100
Hz
250V
1
kHz
1.00
V
10
kHz
19.00V
10
kHz
IIO.OV
10
kHz
19.00V
50
kHz
DISPLAY
READING
247
to
253
2.
On
the
UUT,
select
the
AC mA/A
function,
200
IJ.A
range.
3.
Set the
DMM
Calibrator
for
an output
level
of
190.0
fxA
at
a
frequency
of
1
kHz.
Verify that
the display
reading
is
between
"187.9"
and
"192.1".
4-46,
10A
Current Test
(801
OA
Only)
447.
Use
the
following
procedure
to verify
the
proper
operation
and
accuracy of
the
lOA Range
function of
the
8010A:
1.
Connect
the
HI
output of
the
DMM
Calibrator
to
the
1
OA
input terminal
on
the
UUT.
Connect
the
LO
output of
the
DMM
Calibrator
to the
COMMON
input terminal
on
the
UUT.
Table
4-6.
DC mA
Test
20
mA
200
mA
2000
mA
UUT
INPUT
DISPLAY
READINGS
190.0
mA
1.900
mA
19.00
mA
190.0
mA
1900
mA
189.3 to 190.7
1.893
to
1.907
18.93
to
19.07
189.3
to
190.7
1893
to
1907