12 Error Messages

Errors occurring during operating time are shown on the display and can be deleted with the key Continue. If
the fault persists, the message is repeated and the furnace will work when only when the fault has been
remedied. Should you have any questions with regard to recurrent faults you can call our Hotline no.
07761/562222.
It may be necessary to open the furnace for checking and remedying a fault. Should this be the case, please
heed the safety instructions described in the section "Safety advice"
The following error messages are shown in the display:
Vacuum level of 30 % not reached within 15 sec.
Firing temperature exceeded by 20 °C.
Heating-up temperature exceeded by 30 °C.
Temperature sensor out of order or interrupted.
End switch for lower lift position not reached or out of order.
Fault in the ventilator.
Lift out of order.
Lift blocked.
Power failure lasting more than 10 seconds.
Recover, power failure of less than 10 seconds.
Muffle or fuse out of order.
Program memory full.