B) With high-performance lamp
The adjustment of the microscope and the microscopic image is
done in the same way as described for the halogen lamp
⚪ Switch to the high-performance lamp: switch on the desired
high-performance lamp with corresponding power supply
⚫ Redirect the light with rotating prism into the microscope for
transmitted-light
⚫ On mirror house 4, turn the exciter filter to the white dot. On
mirror house 2, set the exciter filter to red dot
⚫ Set sliding mirror for high-performance lamp. Engage the
neutral density filters for transmitted-light when needed
⚫ Evenly illuminate the field with the high-performance lamp
collector
Immersion darkfield condenser:
N.A.: 1.18 -1.42 oil
Free working distance: 0.38mm
slide thickness: 1.1 ± 0.1mm
Use: for all objectives from 40x to 100x
A cardioid condenser, fully suitable for fluorescence studies
Dry darkfield condenser:
N.A.: 0.7-0.9
Free working distance: 4.8mm
Slide thickness: 1.1 ± 0.1mm
Use: In combination with the condenser turret, can work with 25x
and 40x objectives
An aperture insert is used in the condenser when working with
the Apo 25x and Plan 40x objectives
The condenser is fully suitable for fluorescence studies