! IMPORTANT ! please read
Most common reasons for miss readings
Product temperature out of application range
Material below 0°C resp. above +40°C (32 to 104 °F) may
cause faulty measurements. The storage of cold material in a
warm storage area usually creates condensed water which
may lead to major measuring errors.
Not adjusted material under test
Let
your humimeter
temperature of the material for approx. half an hour.
A very high temperature difference has a negative effect on
the stability of the measurement results.
Wrong calibration curve
Before you measure your sample, double check the correct
selection of the calibration curve.
Wrong filling quantity
Fill in exactly the right weight (± 0.01kg) of wood chips in the
measuring chamber.
Wet or mouldy material
Frozen measuring material
Device maintenance instructions
To provide a long life of your device please does not expose it to
strong mechanical loads or heat e.g. dropping it or direct sunlight
exposure. Clean your device using a dry cloth. The measuring
chamber needs to be cleaned with a dry and soft brush.
Any kind of wet cleaning damages the device. The instrument is
not rainproof. Keep it in dry areas. When the device is not used for
a longer period (6 months) or when the batteries are empty, they
should be removed to prevent a leakage of the battery acid.
page 12