Download Print this page

Nagano Keiki KH41 Manual

Process applications pressure transmitter

Advertisement

Quick Links

B03-02-J
Cat.No.
KH41・ 43, KD41・ 43
Process Applications
Pressure Transmitter
Outline
On-site type pressure transmitter for process use.
Since it is for general applications and proof
pressure explosion-protected use, select the type
corresponding to the working environment.
Features
Since the sensor uses a semiconductor strain
gauge, durability and stability are excellent.
Sinse the wetted parts use a high
corrosion-resistant metal diaphragm, corrosion
resistance is excellent and use with a wide range
of gases and liquids is possible.
EMI countermeasures have been taken so that
transceivers and other electronic devices are not
affected.
Model configuration
General process use
KH41
KH43
Specifications
Fluid:
  Gas, liquid
   *The diaphragm type is available, when the viscosity of
    fluid to be measured is high. Please contact us.
Operating condition:
  KH41, 43 (General process use)
   Under the normal condition,where there is no
inflammable gas or liquid which cause the ignition or
explosion.
  KD41, 43 (Explosion-proof construction)
   Hazardous area
For details, please refer to the description of the
proof pressure explosion-protected construction.
Type:
  Direct mounting, surface mounting, pipe mounting
Connection:
  Rc 1/2, Rc 1/4, G3/8B (Diaphragm only)
Wetted parts material:
  KH41  SCS14 + SUS316
  KH43  SUS316 + Co-Ni alloy
  KD41  SCS14 + SUS316 + SUS304
  KD43  SCS14 + Co-Ni alloy
Pressure range:
   -5 to +5→-50 to +50kPa
   -0.1 to 0→-0.1 to 2MPa
   0 to 5kPa→0 to 50MPa
Power source:
  12 to 32V DC (Standard 24V DC)
Output:
  4 to 20mA DC (2 wire system)
Load resistance:
  600Ω Maximum (24V DC)
Withstand voltage:
  250V AC, 1 minutes
Explosion-proof use (Explosion-proof construction)
KD41 (d2G4)
KD43 (d2G4)
KH41
Withstand voltage:
  250V AC, 1 minutes
Allowable maximum pressure:
  150% of pressure range
Operating temperature range:
  KH41, 43 -30 to 80℃ (No freezing or condensation)
  KD41, 43 -20 to 60℃ (No freezing or condensation)
Storage temperature range:
  KH41, 43 -30 to 80℃ (No freezing or condensation)
  KD41, 43 -30 to 70℃ (No freezing or condensation)
Operating humidity range:
  30 to 95%RH (No condensation)
Accuracy:
  ±0.25%F.S. or ±0.5%F.S.
  (Includes Linearity, hysteresis, repeatability)
Temperature coefficient:
  ±0.025%F.S./℃ (Zero)
  ±0.025%F.S./℃ (Span)
  (-20 to 70℃)
Outlet for electric wire:
  KH41, 43 PF1/2
  KD41, 43 Proofpressure gasket type, conduit type
          (Adjustable cable diameter φ7 to φ12)
Case material:
  ADC12
Protection:
  K□41 Splash-proof (IP54)
  K□43 Jet-proof (IP55)
      However, splash-proof in case of the pressure
range 2MPa or lower (IP54)
Weight:
  Approx. 0.85kg to 5.8kg
KH43
KD41, KD43

Advertisement

loading

Summary of Contents for Nagano Keiki KH41

  • Page 1 Operating condition: Operating temperature range:   KH41, 43 (General process use)   KH41, 43 -30 to 80℃ (No freezing or condensation)    Under the normal condition,where there is no   KD41, 43 -20 to 60℃ (No freezing or condensation) inflammable gas or liquid which cause the ignition or Storage temperature range: explosion.
  • Page 2: Explosion-Proof

    KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Wiring + + K□41 Power source 12-32V DC K□43 − Receiver − + − (Example: GC95 Digital meter relay, etc.) Load resistance range 1000 Load resistance (Ω) Load resistance tolerance 10 12...
  • Page 3: Surface Mounting

    KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Dimensions 1 Unit: mm KH41 Pressure Transmitter Direct mounting (Weight: Approx. 1.8kg) 2B Pipe mounting 70 ±0.3 4-M5 Depth 8 (Back) (41) (Mounting hole) PF1/2 (Outlet for electric wire) PF1/2 (Outlet for electric wire) 2-φ10 Hole...
  • Page 4 KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Dimensions 2 Unit: mm KD41, 43 Explosion-proof construction Pressure Transmitter 70±0.3 70±0.3 (151) 4-M5 Mounting hole (151) 4-M5 Mounting hole (Depth 8) (Depth 8) Earth terminal (M4) Earth terminal φ31 Conduit type...
  • Page 5: Model Number Configuration

    KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Model number configuration 1 Please specify the model number, each specs and the range for ordering. Model K H 4 1 Process Applications ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨...
  • Page 6 KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Model number configuration 2 Please specify the model number, each specs and the range for ordering. Model K H 4 3 Process Applications ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨...
  • Page 7 KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Model number configuration 3 Please specify the model number, each specs and the range for ordering. Model K D 4 1 Process Applications ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨...
  • Page 8 KH41 ・43, KD41 ・43 Process Applications Pressure Transmitter Model number configuration 4 Please specify the model number, each specs and the range for ordering. Model K D 4 3 Process Applications ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨...

This manual is also suitable for:

Kd41Kd43Kh43