Page 1
TD830 E LECT RO SURGICAL UNIT 698257 E-IM54f...
Page 2
(view from rear of unit). For further information visit www.eschmann.co.uk All correspondence relating to the after sales service of Eschmann Equipment to be addressed to : UK Customers Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
TD830 ELECTROSURGICAL UNIT CONTENTS 5. OPERATION 1. PRELIMINARY INFORMATION Preliminary information ......4 Routine check ......16 Connection of accessories .
These Instructions for Use should be referred to for mains supply. A complete systems check must be carried details of the TD830 Electrosurgical Unit, REF 83-256-01, out before using the Electrosurgery Unit (see Sections 5.10 83-257-02, 83-258-03, 83-259-04, 83-260-05, 83-261-01, to 5.17).
ELECTROSURGICAL UNIT Output (bipolar) 2. 0 TECHNICAL DATA Peak GENERAL Open Crest Circuit The TD830 Electrosurgical Unit (or Surgical Diathermy Symbol/Function Power factor Voltage Unit) is classified as ‘HF surgical equipment’ Coagulation - ‘HF surgical equipment’ is defined as, “Medical...
Pinpoint coag (green) Flammable gases Spray coag (bright green) The TD830 equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air or Bipolar: with Oxygen or Nitrous Oxide.
TD830 ELECTROSURGICAL UNIT DUTY CYCLE The symbols adjacent to the mains ‘on/off’ switch The duty cycle rating of 10 seconds ‘on’, 30 seconds ‘off’, (42) indicate the ON and OFF positions respectively. specified on the serial plate as 10s:40s, indicates that the equipment can be used at full output power in any mode for 10 seconds with a 30 seconds rest period.
ALARM SYMBOLS - Indicates the ‘Plate continuity monitor’ alarm has been activated. - Indicates the ‘Plate attachment monitor’ alarm has been activated. - Indicates the ‘Patient earth monitor’ alarm has been activated. - Indicates the ‘Plate voltage monitor’ alarm has been activated.
DO NOT: ♦ ♦ Use only Eschmann accessories, in particular active Do not use uninsulated forceps, monopolar or bipolar. and plate electrodes and cables which should ♦ Do not place monitoring electrodes close to the preferably be no more than 3 metres in length.
If such an earth path is ♦ Contact the hospital electronics engineer or Eschmann present, the alarm will operate until the path has been After Sales Service Department if the equipment fails removed and the unit reset by switching ‘off’...
4.11 The equipment is designed to be used with the output and can be considered as USER 3. Eschmann range of active, plate and bipolar cables. The front panel is colour coded, PALE BLUE for bipolar, OPERATING MODES / DISPLAY OPTION YELLOW for monopolar cut, blend and specialist cut, and BLUE for monopolar coag.
♦ Do not apply the plate electrode adjacent to ECG ELECTROSURGICAL TECHNIQUES electrodes and cables, or to an area where fluids could pool. General ♦ With babies and small children the plate electrode 4.14 When using electrosurgery, it must always be should always be applied to the trunk section of the remembered that its purpose is to cut, destroy or modify body to give maximum contact with the skin.
4.34 If, during a surgical procedure, the output Active Cables appears to drop or the electrosurgical performance 4.25 Only active cables supplied by Eschmann should decreases, stop and check the plate electrode for be used. This is of particular importance, as cables with...
(References ¹ and ²) suggest that an implanted pacemaker affecting the TD830, it is good practise, in common with all should be unaffected, or only revert to fixed-rate operation, electrosurgical units, to keep them as far as is reasonably possible from the surgical site.
S-Lead [REF 83-139-41 (Olympus 9mm), or REF 83-139-49 (Pentax 8mm)] connected to the 4.54 This electrosurgical unit has dual monopolar outputs S-Lead socket (29, Fig 12) of the TD830 (first come first served) and simultaneous bipolar output. Electrosurgical unit and the flexible endoscope.
If in doubt, discard suspect component and fingerswitch sockets (25) can cause damage substitute with one of known function, to the TD830 if the output is energized by a continuity and condition. footswitch. ♦ A single active cable with 8 mm plug (footswitch ♦...
(42) will come ‘on’ and the front (Fig. 7). panel displays (indicators and digital) will illuminate. ♦ For foot control, use one of the range of Eschmann 5.12 The unit may be in monopolar standby mode, if it is footswitches together with a non-switching active monopolar displays will not be illuminated and monopolar handle.
USER 1 and/or USER 2 selection covered by adhesive. With cut and coag power levels 5.19 The TD830 can be used by two monopolar operators both adjusted to zero (for safety) the PEM alarm USER 1 and USER 2 independently of each other but not should operate when a USER 1 footswitch or simultaneously.
♦ The plate electrode is correctly positioned, and is in contact with the patient. 5.32 The Eschmann twin footswitch (Fig. 4) has an angled platform with two light-action switches for cut and coag. ♦ The plate electrode cable is intact, and is plugged into Colour coded bars provide foot location and protection unit plate electrode socket (30).
especially flexible endoscopes. Most urological WARNINGS endoscopes are suitable for spray coag. If in doubt check Footswitches should be used with care at that the voltage rating of the laparoscopic or endoscopic all times and should always be under control instrument is at least as high as the maximum spray coag open circuit voltage of this electrosurgical unit (see of the surgeon or designated assistant, as...
The pneumatic type has a very light action and is characteristics determine the actual power delivered to the suited to procedures where repetitive actuation is needed. tissue for a given control setting. The TD830 electrosurgical unit has a power curve that rises at the low resistance Note: end, this ensures that adequate power is available for ♦...
5.55 The output control settings required for a particular 5.57 If more than one output demand is received at the surgical procedure are also dependent on a number of same time (such as two footswitches or a footswitch and a external factors (type of tissue, speed at which the surgeon fingerswitch button being pressed) the unit will produce works, type of electrode in use, area of forcep tips in...
Page 23
Note: An activated PVM alarm will automatically be reset by releasing a depressed footswitch or fingerswitch. If none of the above remedies stop the alarm activated, or the (!) alarm is activated contact Eschmann Equipment and do not use the unit. E-IM54f P23/37...
Table 1 Power Output Selection Guide (for ‘typical’ power display range) Procedure Mode Condition Setting CAUTION Do not use a higher power setting than is necessary General Normal cut Dry field 70 to 90 Pinpoint coag Dry field 25 to 45 Blend Dry field 45 to 70...
TD830 ELECTROSURGICAL UNIT MONOPOLAR CUT DIAGRAMS MONOP OLA R C UT (P OW E R v LOA D ) H a lf o u tp u t co n trol s e ttin g Full ou tpu t con tro l s e ttin g...
Page 26
MONOPOLAR BLEND DIAGRAMS MONOP OLA R B LE ND (P OW ER v LOA D ) H alf ou tp ut co ntrol s etting Fu ll o utpu t co n tro l s e ttin g 1000 1500 2000 Load resistance (ohms) MONOPOLAR BLEND (POWER v CONTROL)
Page 27
TD830 ELECTROSURGICAL UNIT MONOPOLAR SPECIALIST CUT DIAGRAMS MONOP OLA R S P E C IA LIS T C UT (P OW E R v LOA D ) H a lf ou tp u t co ntro l s e tting...
Page 28
MONOPOLAR PINPOINT COAG DIAGRAMS MONOP OLA R P INP OINT C OA G (P OW E R v LOA D ) H alf ou tpu t con tro l s e ttin g Fu ll o u tp ut co n tro l s etting 1000 1500 2000...
Page 29
TD830 ELECTROSURGICAL UNIT MONOPOLAR SPRAY COAG DIAGRAMS M O NO P O LA R S P RA Y C OA G (P OW E R v L O A D ) H a lf o u tp u t co n tro l s e ttin g...
Page 30
BIPOLAR MACRO DIAGRAMS BIPOLAR MAC RO (POWER v LOAD) Macro - H alf output control s etting Macro - Full output control s etting 1000 Load resistance (ohm s) BIPOLAR MACRO (POWER v CONTROL) BIPOLAR MACRO (VOLTAGE v CONTROL) Load resistance 100 (ohms) Open circuit Output control setting (numerical value up to 10) Output control setting (numerical value up to 10)
Page 31
TD830 ELECTROSURGICAL UNIT BIPOLAR MICRO DIAGRAMS B IPOLA R MIC RO (P OW ER v LOAD ) Micro - H alf output control s etting Micro - Full output control s etting 1000 Load resistance (ohm s) BIPOLAR MICRO (POWER v CONTROL)
Data). If repeated blowing of a fuse(s) occurs, either when wiped dry. the unit is switched on or during use, call the Eschmann ♦ Do not use excessive amounts of liquid which could After Sales Service Department (see inside front cover).
TD830 ELECTROSURGICAL UNIT 7.0 ACCESSORIES The accessories in the list below and illustrated where referenced, are all recommended for use with the TD830 electrosurgical unit. (REF = catalogue number) Description Illustration No. Description Illustration No. Active electrode silicone cables Colposcopy electrodes (LLETZ) cone biopsy...
Need help?
Do you have a question about the TD830 and is the answer not in the manual?
Questions and answers