Page 2
Calibration Statement The Rigel Multi-Flo Infusion Device Analyser is fully calibrated and found to be within the specified performance and accuracy at the time of production. The Seaward Group provides its products through a variety of channels; therefore it may be possible that the calibration date on the provided certificate may not represent the actual date of first use.
Page 4
Disposal of old product The Rigel Multi-Flo has been designed and manufactured with high quality materials and components, which can be recycled and reused. When this symbol is attached to a product it means the product is covered by the European Directive 2002/96/EC.
Page 5
Peterlee, County Durham, SR8 2SW, England As the manufacturer of the apparatus listed, declare under our sole responsibility that the product: Rigel Multi-Flo Infusion Device Analyser To which the declaration relates are in conformity with the relevant clauses of the following standard: BS EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements.
Setting the Time/Date ..................8 Displaying the Multi-Flo Information ..............9 Analysing an Infusion Device................ 10 Connecting an Infusion Device to the Multi-Flo ..........10 Priming the Channels ..................10 Patient Controlled Analgesia (PCA) test............11 Occlusion test ....................12 Flow Rate/Volume test..................
Measuring flow rates, volume and pressure, the Multi-Flo is available in 1, 2 and 4 channel configuration which can be upgraded in the future to include additional channels up to a maximum of 4 channels. The Multi-Flo will ensure it meets your current and future requirements.
Carefully unpack all items from the box and ensure the following items are included: Rigel Multi-Flo Infusion Device Analyser Mains Power Lead Multi-Flo Quick Start Guide Utilities Disc USB Bluetooth Adaptor 3 | Page Rigel Medical Multi-Flo user Manual V1.0...
Safety - Ensure that only accessories supplied by the manufacturer or accessories that meet the manufacturer’s specification are used. Safety - Where safe operation of the Multi-Flo is no longer possible it should be immediately shut down and secured to prevent accidental operation.
Page 10
It must be assumed that safe operation is no longer possible: if the instrument or leads show any sign of damage or the instrument does not function or after long periods of storage under adverse environmental conditions. 5 | Page Rigel Medical Multi-Flo user Manual V1.0...
Function keys F1 - F4 Start button Rotary encoder Stop/End button. Channel inlet connectors. Channel drain connectors. Folding legs Type A USB connection Type B USB connection IEC mains power lead connection Auxiliary output connection 6 | Page Rigel Medical Multi-Flo user Manual V1.0...
Ensure that the Multi-Flo is operated with the distilled or de-ionised water only. The Rigel Multi-Flo has two legs on the front base of the unit. These are designed to raise the front of the unit to improve the viewing angle of the colour graphic display.
Alternatively, use a USB keyboard to enter data. Press 12/24 to toggle between 12 and 24 hour clock mode. Press Format to toggle between DDMMYYYY and MMDDYYYY date format. Press to save the changes, then to exit. 8 | Page Rigel Medical Multi-Flo user Manual V1.0...
Press Details to view the firmware version, hardware version and serial no. for the Multi- Flo unit and each individual installed channel. Press Calibration to view the calibration data of the installed channels. Press to view the Service and Calibration contact information. 9 | Page Rigel Medical Multi-Flo user Manual V1.0...
Priming the Channels Upon power-up the Multi-Flo is in automatic priming mode however, should priming be required at any other stage, select the PRIME from the main menu;...
BOLUS. Therefore an incorrect setting of the basal flow rate will lead to an inaccurate BOLUS detection. Press to save and advance to the PCA test summary screen, or to exit without saving. 11 | Page Rigel Medical Multi-Flo user Manual V1.0...
From the main menu, select Manual mode. This will take you to the channel summary screen. Highlight the required channel and press SET UP. This will take you to the Occlusion Setup screen. 12 | Page Rigel Medical Multi-Flo user Manual V1.0...
Page 18
The automatic pump setting refers to infusion devices that do provide an automatic pressure back-off function. The Multi-Flo will detect the occlusion alarm when the line pressure decreases after alarm. Press...
Flow Rate/Volume test The Rigel Multi-Flo is capable of measuring the instantaneous flow at a resolution of 10µl/hr. In addition, the flow rate can be viewed based over an average period (user selectable) as well as detecting peak and minimal flow rates.
The graph can switch between instantaneous flow and volume by pressing the TYPE button. Draining the Channels From the main menu, select DRAIN. This will take you to the Drain function screen. Either select the individual channel to drain or select Drain All. 15 | Page Rigel Medical Multi-Flo user Manual V1.0...
Page 21
Once the fluid has been drained, press to stop the draining process. Warning; Once the fluid is drained, do not leave the drain pump facility running longer than necessary. 16 | Page Rigel Medical Multi-Flo user Manual V1.0...
Avoid using solvents and abrasive scouring agents to clean the external case of the Rigel Multi-Flo. If the Multi-Flo is subject to liquid ingress in a manner other than intended, the unit should be returned for repair, stating clearly the cause for repair.
Accessories Optional Accessories The 1 and 2 channel Multi-Flo configurations are field upgradeable to a maximum of 4 channels. USB keyboard USB download lead Med-eBase PC download and remote control software Replacement Spare Parts 44B122 - IEC mains lead 27B044 - Mains fuse, 20 x 5mm T3.15A 250V...
0.1 ml to 100 ml Measuring Range 0.5 ml to 100 ml Accuracy ± 1% of the reading Max. resolution 0.01 ml Basal flow rate 1 ml to 30 ml/h Pressure Max. 2500 mmHg 20 | Page Rigel Medical Multi-Flo user Manual V1.0...
USB A Display LCD colour graphic display ¼” VGA Environmental Conditions The Rigel Multi-Flo has been designed to perform tests and measurements in a dry environment. Maximum barometric elevation for making measurements is 2000m. Protective system IP40 according to IEC 60529.
CalibrationHouse Tel: +44 (0) 191 587 8739 11 Bracken Hill Fax: +44 (0) 191 518 4666 South West Industrial Estate Email: info@calibrationhouse.com Peterlee, County Durham SR8 2SW, United Kingdom Part of 22 | Page Rigel Medical Multi-Flo user Manual V1.0...