Low Battery Alarm – The message "Low Batt" is displayed on the screen and the battery level
indicator will turn RED. When this warning is displayed, the unit will shut off automatically after
a few minutes of continued use, but BP may not be measurable even though SpO2 is still
functioning. When this occurs, the AC/DC adapter should be plugged in to prevent automatic
shutdown, or the unit should be taken out of service and recharged.
The system alarms may also be silenced by touching anywhere on the screen, but take care not to touch
a button unintentionally when canceling an alarm condition.
BP measurement
not achieved
O2 Sensor Alarm
Low Batt Alarm