The Levelmaster commands are based on the following syntax:
Capital letters are at the beginning of certain data fields
Small letters stand for data fields
All commands are terminated with "<cr>" (carriage return)
All commands start with "Uuu", whereby "uu" stands for the address (00-31)
"*" can be used as a joker for any position in the address. The sensor always converts this in
its address. In case of more than one sensor, the joker must not be used, because otherwise
several slaves will answer
Commands that modify the instrument return the command with "OK". "EE-ERROR" replaces
"OK" if there was a problem changing the configuration
Report Level (and Temperature)
Request:
Parameter
Report Level (and Temperature)
Response:
Parameter
Report Level (and Temperature)
Level in inches is repeated for sensors with 2 levels. PV value is transmitted as first measured value,
SV as second measured value.
PV, SV, TV and QV can be adjusted via the sensor DTM.
Report Unit Number
Request:
Parameter
Report Unit Number
92
Configurable Values
1200, 2400, 4800, 9600, 19200
1
7, 8
None, Odd, Even
1, 2
32
Length
4 characters ASCII
Length
24 characters ASCII
Length
5 characters ASCII
SITRANS LG270 - Operating Instructions
Default Value
9600
1
8
None
1
32
Code/Data
Uuu?
Code/Data
UuuDlll.llFtttEeeeeWwwww
uu = Address
lll.ll = PV in inches
ttt = Temperature in Fahrenheit
eeee = Error number (0 no error, 1
level data not readable)
wwww = Warning number (0 no
warning)
Code/Data
U**N?
PBD-51041066