Humidity
Vibration
Altitude
Electromagnetic Compliance Specifications
Appendix A
Operating Humidity Range:
•
Up to 95% RH from 0°C to +40°C (non-condensing).
•
Up to 65% RH from +40°C to +55°C (non-condensing).
Storage Humidity Range:
•
Up to 95% RH from 0°C to +55°C (non-condensing).
•
Up to 90% RH from +55°C to +75°C (non-condensing).
Shock
End Use Handling:
Half sine waveform, <3 msec duration, ∆v = 160 cm/sec minimum.
•
Transportation:
Trapezoidal waveform, ∆v = 605 cm/sec, 30 g minimum.
•
Operating and Functional:
•
5 to 500 Hz, 0.0001 g
Survival, Swept Sine:
•
5 to 500 Hz Resonance Search, 5 minute dwell on resonances at 0.5g.
Survival, Random:
•
0.015 g
Up to 3000m.
The HP E8400A mainframe conforms to the following EMC product
specifications:
•
CISPR 11:1990/EN55011 (1991): Group 1 Class A
•
EN61000-3-2:1995 Class A
•
EN61000-3-3:1995
•
EN50082-1:1992
•
IEC 801-2:1991: 4kV CD, 8kVAD
•
IEC 801-3:1984: 3 V/m
•
IEC 801-4:1988: 1kV Power Line, 0.5kV Signal Lines
•
ENV50141:1993/prEN50082-1 (1995): 3 Vrms
•
ENV50142:1994/prEN50082-1 (1995): 1 kV CM, 0.5 kV DM
•
EN61000-4-8: 1993/prEN50082-1 (1995): 3 A/m
•
EN61000-4-11:1994/prEN50082-1 (1995): 30 %, 10 ms 60 %, 100ms
2
/Hz Spectral Density
2
/Hz Spectral Density.
Specifications
55
Need help?
Do you have a question about the E8400A C-Size VXI Mainframe and is the answer not in the manual?
Questions and answers