Invacare® Pegasus Series
Flash code
Fault
5
Brake failure
6
No neutral position when
switching scooter on.
7
Fault in speed
potentiometer
8
Motor voltage error
9
Miscellaneous internal
fault
10
Push/freewheel mode
error
58
Consequence for the
scooter
Comments
Stops driving
•
•
Stops driving
•
•
Stops driving
•
•
Stops driving
•
Stops driving
•
Stops moving
•
Ensure that the disengaging lever is in the engaged
position.
There is a defect in the braking coil or in the cabling.
Check the magnetic brake and cabling for open
or short-circuited circuitry. Contact your Invacare
provider.
Drive lever is not in neutral when the keyswitch was
turned. Put the drive lever in neutral, turn the power
off and then turn on again.
It may be necessary to replace the drive lever. Contact
your Invacare provider.
The drive lever controls could be faulty or
incorrectly connected. Check the cabling for open or
short-circuited circuitry.
Potentiometer is not correctly adjusted and must be
replaced. Contact your Invacare provider.
The motor or its cabling is defective. Check the cabling
for open or short-circuited circuitry.
Contact your Invacare provider.
The scooter has exceeded the permissible maximum
speed during pushing or freewheeling. Switch the
electronics system off and on again.
1638629-A