LABOPORT ® Chemically-resistant Laboratory Pumps with Modular Accessories Operating Instructions for Laboratory Pumps N810, N820, N840, N842 and for LABOPORT Systems SR..., SH…, SC…, SCC… Operating Instructions Vacuum-Controller NC800 / NBC800 KNF 121722-121728 02/10 Translation of original Operating Instructions, english...
Page 2
KNF Neuberger UK Ltd Ave 2, Station Lane Ind Est, Witney, Oxfordshire, OX28 4FA, UK Tel: +44 (0) 1993 778373 Fax: +44 (0)1993 775148 E-Mail: info@knf.co.uk www.knf.co.uk...
An activity to be carried out (a step) is specified here. 1. The first step of an activity to be carried out is specified here. Additional, consecutively numbered steps follow. This symbol refers to important information. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
The pump/the system must not be used to create vacuum and overpressure simultaneously. An overpressure must not be applied to the suction side of the pump/ the system. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
If necessary, consider any external sources of energy, such as radiation, that may add heat to the medium. In case of doubt, consult the KNF customer service. Environmental protection Store all replacement parts in a protected manner and dispose of them properly in accordance with the applicable environmental protection regulations.
Page 7
II the pollution degree 2 Customer service and Only have repairs to the pump/system carried out by the KNF repairs Customer Service responsible. Only authorized personnel should open those parts of the housing that contain live electrical parts.
15 Holder for Separator 16 Support for vacuum control- 17 Baseplate 18 Vacuum pump 19 On/off switch of pump Fig. 2: Full expanded LABOPORT System SCC..., for example pump model N 840.3 FT.18 (symbolic) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 9
Electrical supply unit 1 vacuum controller 2 vacuum controllers Tab. 2 Each pump and all systems can be expanded modularly up to a full system SCC… (see chapter 7.7, page 28). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Maximum permitted mains +/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 1.25 (2x) T [A] Protection class motor IP44 Tab. 4 * Liters in standard state (1,013 mbar) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 11
Maximum permitted mains +/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 3.15 (2x) T [A] Protection class motor IP44 Tab. 5 * Liters in standard state (1,013 mbar) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 12
+/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 3.15 (2x) T [A] Protection class motor IP44 Tab. 6 * Liters in standard state (1,013 mbar)N 810.3 FT.18 Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 13
Maximum permitted mains +/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 1.25 (2x) T [A] Protection class motor IP44 Tab. 7 *Liters in standard state (1,013 mbar) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 14
Maximum permitted mains +/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 3.15 (2x) T [A] Protection class motor IP44 Tab. 8 *Liters in standard state (1,013 mbar) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 15
Maximum permitted mains +/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 3.15 (2x) T [A] Protection class motor IP44 Tab. 9 *Liters in standard state (1,013 mbar) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 16
Maximum permitted mains +/- 10 % +/- 10 % +/- 10 % voltage fluctuation Fuse pump 3.15 (2x) T [A] Protection class motor IP44 Tab. 10 *Liters in standard state (1,013 mbar) Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
* includes the power consumption of the whole system (inclusive pump and vacuum controller); for pumps N 840.1.2 FT.18 and N 842.3 FT.18 power consumption raises by approx. 50 W. 5.3. Vacuum Controller See Operating Manual of the vacuum controller. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Operating parameters of coolant (only for high performance condensers) Permissible pressure [bar g] Permissible temperature - 15 °C to + 20 °C Condenser connectors For tube 8mm ID Tab. 13 Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
(1). The transfer chamber (3) is hermetically separated from the pump drive (7) by the diaphragm. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
On the suction side the separator collects particular matter and droplets. This protects the pump from contamination. The separator is made of a specially treated glass and features implosion protection. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
The condensate is collected in a glass flask. The flask is attached to the condenser flange with a clamp. The condensation tempera- ture is maintained by running cold water or recirculated coolant through the condenser. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
If one of the two vacuum values falls below the hysteresis range, the pump will switch back on. Refer to the vacuum controller’s operating instructions for more information. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
3. Lay the suction and pressure line at a downward angle to prevent condensate from running into the pump. 4. Insert the power cable’s plug into a properly installed shockproof socket. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
3. Place the glass separators into the holders. Laying hoses: depending on system configuration see chapter 7.7.1 (system SR), 7.7.2 (system SH), 7.7.3 (system SC), 7.7.4 (system SCC) or 7.7.5 (retrofitting from system SC to system SCC). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Allen key 4 mm Allen key 5 mm Tab. 16 Mounting the electrical supply unit 1. Slide the electrical supply unit (see fig. 8/4) with both tightening nuts into the baseplate’s lateral slotted bar. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 26
8. Make sure power switches on the pump and on the vacuum controller(s) are switched off. 9. Insert the electrical supply unit`s system power cord plug into a shockproof socket. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 27
1. Connect the coolant valve to the coolant supply. 2. Insert the coolant valve’s plug into the vacuum controller’s VV1 socket (see fig. 10/3). See fig. 16, page 29 for coolant inlet and outlet. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Fig. 13: Mounting the gas ballast 7.7. Mounting of Systems The displayed system tube configuration is for two-headed pumps only. Contact KNF Service for information about laying tubes for one-headed pumps and pumps with aluminum heads (“A” in the type designation). 7.7.1. System SR The system SR consists of: ...
Make sure that the coolant valve is mounted Fig. 16: Coolant supply of high between the coolant supply and the coolant inlet performance condenser port of high performance condenser. WARNING 1 Coolant inlet 2 Coolant outlet Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
WARNING 6. Wire and connect with each other the electrical supply unit and the vacuum controller (see chapter 7.5, page 21). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
WARNING 6. Wire and connect with each other the electrical supply unit and the vacuum controller (see chapter 7.5, page 21). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
6. Lay hoses for system (see fig. 19). Fig. 19: Tubing system SCC 7. Wire and connect with each other the electrical supply unit and the vacuum controller (see chapter 7.5, page 25). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Make sure the pumps are used properly (see chapter 2.1, page 5). Make sure the pumps are not used improperly (see chapter 2.2, page 5). Observe the safety precautions (see chapter 3, page 6). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 34
Attempt work on the pump or system only if the pump/system is separated from mains power. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
1. Loosen the both allen screws at the foot of the high perform- ance condenser. 2. Remove the high performance condenser’s clamp assembly in the support mounting slot. 3. Retighten the allen screws. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
If the gas ballast valve is open the maximum achievable vac- closed uum level is reduced. The gas ballast is adjusted with the button (see fig. 20). open Fig. 20: Gas ballast operating knob Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
1. Carefully remove the spring clamp (fig. 2/10) while supporting the flask (fig. 2/11). 2. Empty and clean the flask – observing safety precautions. 3. Re-connect the flask to the condenser and replace the spring clamp. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Depending on the substance transferred, caustic burns or poisoning are possible. WARNING Wear protective clothing if necessary, e.g. protective gloves. Flush pump before replacing the diaphragm and valve plates/sealings (see chapter 9.2.1, page 37). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 39
4. Screw in the new diaphragm and tighten it by hand. 5. With a two-headed pump: Complete steps 1 through 4 for the second pump head. Fig. 22: Changing structured diaphragm Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 40
9. For two-headed pumps: Refit the pneumatic head connection: Place tube onto the connecting part of the angle fitting, turn angle fitting to a straight position and tighten the union nut. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 41
LABOPORT Pumps and Systems Servicing Final steps 1. Remount the pump to the baseplate (if applicable). 2. Reconnect suction and pressure line to the pump. 3. Reconnect the pump to the electricity supply. Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Remove the jamming parts and particles. Head parts are soiled. Clean head components. Diaphragm or valve Replace diaphragm and valve plates/sealings, plates/sealings are worn. (see chapter 9.3, page 38). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Solid particles in the pump vent Clean pump vent valve. valve Condensate in pump head Let the pumpe run to the ultimate vacuum (Drying Mode, see Operating Manual of Vacuum Controller). Tab. 30 Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
Page 44
Vacuum Controller shows unrealistic values Cause Fault remedy The factory-set pressure Contact KNF (see page 47). equalization has changed. Sensor is defective Contact KNF (see page 47). Tab. 34 Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
10.4. Fault cannot be rectified If you are unable to determine any of the specified causes, send the pump to KNF Customer Service (see page 47 for the address). 1. Flush the pump to free the pump head of dangerous or ag- gressive gases (see chapter 9.2.1, page 37).
Vacuum chemical tubing for pneu- X 028187 matic connection** (yard ware***) Tab. 37 * For retrofitting from system SC to SCC ** Material: Norprene A60G *** Please specify the required length (in whole meters). Translation of original Operating Instructions, english, KNF 121211-121345 01/10...
LABOPORT Pumps and Systems Decontamination Declaration 12. Decontamination Declaration The condition for the repair of a pump by KNF is the certifica- tion of the customer on the transferred media and on the cleaning of the pump (decontamination declaration). Copy this page.
Page 49
INNOVATIVE TECHNOLOGY WORLDWIDE Operating Manual NC 800/NBC 800 Vacuum Controller for Laboratories Carefully study the Operating Instructions and observe at all times the relevant instructions to avoid dangerous situations. KNF 121205-121650 02/10 Translation of original Operating Instructions, english...
Page 50
9 Cleaning gas connection/Aeration Vacuum System Vacuum Valves 10 Coolant valve (accessory) 4 Vacuum valve 11 Cooling water tubing 5 Venting valve (internal) Fig. 1: Arrangement of vacuum controller in vacuum system Translation of original Operating Instructions, english, KNF 121205-121650 02/10...
Changing of pressure unit see section 6. · Connect Vacuum Controller electri- cally LED for pressure display in torr section 1.4 ½ Changing of pressure unit see » Connect Vacuum Controller pneu- section 6. Translation of original Operating Instructions, english, KNF 121205-121650 02/10...
In addi- t [min] tion, for post distillation the pressure setpoint can be lowered precisely Fig. 5: Controlling profile in manual mode: Controlled pressure reduction according to a pre-set time. Translation of original Operating Instructions, english, KNF 121205-121650 02/10...
Press the SET key four times. ½ SET-key: Mode, pressure setpoint and actual pressure are displayed. Changing readings in this order: » Press the STOP key to finish. Pressure setpoint Pressure differential h pressure reduction d Translation of original Operating Instructions, english, KNF 121205-121650 02/10...
Press the STOP key to finish the times. ration. evacuation. Once contact is released from ½ The "mbar" displays functions the UP key the system will eva- Translation of original Operating Instructions, english, KNF 121205-121650 02/10...
Page 55
STOP key however not until the first boiling (and possibly changing the d and t Input post distillation time (min) point has been achieved. values), the AUTO key has to be Translation of original Operating Instructions, english, KNF 121205-121650 02/10...
Tel: +44 (0)1993 778373 Fax: +44 (0)1993 775148 LABOBASE ® Systems E-mail: info@knf.co.uk ³ All workstations together demand www.knf.co.uk a flow rate that exceeds the capa- Translation of original Operating Instructions, english, KNF 121205-121650 02/10...