EN 60601-1-2, Medical Device Directive 93/42/EEC.
These limits are designed to provide reasonable
protection against harmful interference in a
typical medical installation. However, because of
the proliferation of radio-frequency transmitting
equipment and other sources of electrical noise
in the health-care and home environments it is
possible that high levels of such interference due to
close proximity or strength of a source may result in
disruption of performance of this device.
• This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and
used in accordance with these instructions, may
cause harmful interference with other devices in the
vicinity. Disruption or interference may be evidenced
by erratic or incorrect functioning. If this occurs,
the site of use should be surveyed to determine
the source of this disruption, and actions taken to
eliminate the source. The user is encouraged to try
to correct the interference by one of the following
measures:
– Turn equipment in the vicinity off and on to
isolate the offending equipment.
– Reorient or relocate the other receiving device.
– Increase the separation between the interfering
equipment and this equipment.
– If assistance is required, contact your
DePuy Mitek Representative.
• The wireless footswitch, in conjunction with the
DePuy Mitek VAPR VUE generator, has been tested
and found to properly operate in the presence of the
following potentially interfering devices:
– Electrosurgery generators – located on the same
cart as the VAPR VUE generator
– Wi-Fi devices – minimum 3 feet away from
VAPR VUE generator and wireless footswitch
– Bluetooth device – minimum 3 feet away from
VAPR VUE generator and wireless footswitch
– Wireless phones (2.4 GHz) – minimum 3 feet
away from VAPR VUE generator and wireless
footswitch